Method comparison and agreement of a fluorescence immunoassay-based HbA1c test with a routine HPLC method

Oct 2026 · Diabetology International · Vol 17 · 0 citations · 26 references
Medicine

TL;DR

The TRUSTlight FIA showed a small mean bias relative to the HPLC comparator, although the observed limits of agreement were wider than the ± 5% NGSP manufacturer certification criterion, which is wider than the ± 5% NGSP manufacturer certification criterion.

Read PDF

Similar papers

Oct 2026

A-136 Improving HbA1c Result Reportability Using Dual-Method Verification: Bio-Rad D-100 HPLC and Siemens Healthineers DCA Immunoassay

Hemoglobin A1c (HbA1c) is a critical biomarker for diabetes diagnosis and monitoring long-term glycemic control. Separation-based methods, such as the Bio-Rad D-100 ion-exchange HPLC and the Sebia Capillarys® Octa 3 capillary electrophoresis system, provide excellent resolution of hemoglobin species and variants,...

Shu-Ling Fan, W. Gray · 0 citations
Oct 2026

A-147 Performance of an IFCC-Compliant Immunoturbidimetric HbA1c Assay: A Study of Consistency with the Gold Standard

Glycated hemoglobin (HbA1c) reflects average blood glucose over 2–3 months and is the gold standard for diabetes diagnosis and monitoring, as recommended by the American Diabetes Association. High-performance liquid chromatography (HPLC) is the gold standard for HbA1c but is slow and costly. Traditional immunotu...

Cheng-Dong Ji, Yun Li, Qian Chen et al. · 0 citations
Oct 2026

B-059 Precision and Bias Verification of HbA1c by HPLC Following CLSI EP15-A3 in a Hospital Laboratory in Manabí, Ecuador

Glycated Hemoglobin (HbA1c) is a key biomarker for the diagnosis and monitoring of diabetes mellitus, reflecting the average blood glucose over the past 2–3 months. The clinical reliability of the results depends on the proper analytical performance of the method, particularly in terms of precision and bias, whic...

Jorge A. Zambrano, Ana Alcívar, Ivón Howland · 0 citations
Oct 2026

B-299 Analytical performance, variant detection and result reportability of the CAPILLARYS 3 TERA HbA1c assay relative to the Siemens Atellica CH930

Accurate hemoglobin A1c (HbA1c) measurement is key for diabetes diagnosis and monitoring. Many methods exist; however, immunoassay methods may be limited in detection of hemoglobin variants, abnormal fractions, and hemoglobin concentrations that impact result measurement. This study evaluated the analytical perfo...

I. Tarhoni, Kelsey Reschly, M. Edward et al. · 0 citations
Oct 2026

A-180 Measurement of hemoglobin A1c across three methodologies: A comparison in the presence and absence of hemoglobin variants

The D-100, Alinity, and Roche platforms demonstrated high comparability in HbA1c samples without hemoglobin variants and in the presence of variants, particularly elevated HbF, both Alinity and Roche showed modest negative proportional bias relative to D-100 but maintained a strong correlation.

M. Berman, Casey Williamson, P. Cheng et al. · 0 citations
Oct 2026

B-233 Haemoglobin Variants and Their Impact on HbA1c Measurement: Experience with Two Routine Methods

Comparing HbA1c results obtained by PEIT with those from HPLC to evaluate agreement and potential clinical interchangeability in patients with common haemoglobin variants shows acceptable agreement without evidence of substantial systematic or proportional error.

A. Mir, Koushik Biswas · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.