B-233 Haemoglobin Variants and Their Impact on HbA1c Measurement: Experience with Two Routine Methods
TL;DR
Comparing HbA1c results obtained by PEIT with those from HPLC to evaluate agreement and potential clinical interchangeability in patients with common haemoglobin variants shows acceptable agreement without evidence of substantial systematic or proportional error.
Abstract
Haemoglobin A1c (HbA1c) measurement is central to the diagnosis and monitoring of long-term glycemic control in people with diabetes mellitus. High-performance liquid chromatography (HPLC) is widely regarded as the gold standard method, whereas particle-enhanced immunoturbidimetry (PEIT) offers operational simplicity and faster turnaround. The accuracy of HbA1c methods can be affected adversely by the presence of haemoglobin (Hb) variants. This study compared HbA1c results obtained by PEIT with those from HPLC to evaluate agreement and potential clinical interchangeability in patients with common haemoglobin variants. Paired HbA1c values from sixty-seven patients were analysed using HPLC and PEIT. Among the 67 samples analysed, Hb variants were distributed as HbD (n=26), HbE (n=28), increased HbF (n=5), HbS (n=4), increased HbA2 (n=2), and HbC (n=2). Data analysis was undertaken for the entire cohort of 67 samples, as well as separately for the subgroups carrying HbD and HbE variants. Agreement was assessed using Bland–Altman analysis and Passing–Bablok regression. In the entire cohort of 67 samples, Bland–Altman analysis revealed a mean bias of –1.08%, indicating that the test method tends to underestimate HbA1c levels. The 95% limits of agreement suggested that, in individual cases, results may vary by approximately ±3% relative to the reference method. Passing–Bablok regression yielded the equation y = –0.9 + 1.0x, with a slope of 1.0 and an intercept of –0.9. The slope being close to unity indicates good proportional agreement between the two methods. The negative intercept suggests a small constant bias, which is not statistically significant. Overall, the two methods demonstrate acceptable agreement without evidence of substantial systematic or proportional error. A strong correlation was observed between the two methods (r = 0.86, p < 0.0001). In the subgroup HbD variant, Passing-Bablok regression yielded the equation y = –1.187 + 1.07x. The slope being slightly above 1 suggested a tendency toward proportional overestimation by the test method; however, the wide confidence interval (0.596–1.50) indicates substantial uncertainty. The negative intercept suggested a slight constant underestimation. Bland–Altman analysis showed a mean bias of –1.16%, indicating that the test method consistently underestimates HbA1c. The 95% limits of agreement suggested that, for individual patients, results may differ by as much as ±3–3.5% from the reference method. In the subgroup of HbE variant, Bland-Altman analysis revealed a mean bias of –1.06%. The 95% limits of agreement showed that individual measurements may vary by up to nearly –3% on the lower side and +0.6% on the higher side. Passing–Bablok regression yielded a slope of 1.036 and an intercept of –1.422. There is no significant proportional bias between the two methods. However, the intercept suggested the presence of a constant negative bias. Although PEIT correlates strongly with HPLC, it consistently underestimates HbA1c, particularly at higher concentrations. While the negative bias suggests a systematic underestimation, the relatively wide limits of agreement highlight the potential clinical impact, particularly around diagnostic and therapeutic decision cut-offs. Therefore, caution is warranted when interpreting HbA1c results near threshold values.