B-059 Precision and Bias Verification of HbA1c by HPLC Following CLSI EP15-A3 in a Hospital Laboratory in Manabí, Ecuador

Oct 2026 · Clinical Chemistry · 0 citations

Abstract

Glycated Hemoglobin (HbA1c) is a key biomarker for the diagnosis and monitoring of diabetes mellitus, reflecting the average blood glucose over the past 2–3 months. The clinical reliability of the results depends on the proper analytical performance of the method, particularly in terms of precision and bias, which must be verified according to international standards and regulatory criteria. The aim of this study was to assess the precision and bias of the HbA1c method by HPLC following the CLSI EP15-A3 protocol in a hospital laboratory. A Mindray H50P analyzer was used for HPLC to evaluate two levels of external control expressed as % NGSP, following the CLSI EP15-A3 protocol with a 5×5 design (n=25 per level). The results showed low intra-laboratory imprecision (=1.08%) and temporal stability of the method. The observed bias was minimal (0.18–0.53%) and remained well below the maximum allowable bias according to CLIA 2024 (8% of the assigned value: 0.45–0.91 units depending on the level), confirming high accuracy. The HbA1c test by HPLC demonstrated high precision and minimal bias. These findings support the clinical implementation of the method, its integration into internal quality control, and ensure reproducible, reliable results that comply with clinical metrology standards