The D-100, Alinity, and Roche platforms demonstrated high comparability in HbA1c samples without hemoglobin variants and in the presence of variants, particularly elevated HbF, both Alinity and Roche showed modest negative proportional bias relative to D-100 but maintained a strong correlation.
Abstract
Hemoglobin A1c (HbA1c) is a standardized biomarker used in the diagnosis and management of diabetes. Since HbA1c cutoffs are universally applied across clinical laboratories, agreement between analytical platforms is essential. Methodological differences among systems (chromatographic, enzymatic, and immunochemical) may lead to clinically meaningful variability. In addition, developmental and pathological hemoglobin variants, including elevated fetal hemoglobin (HbF), may differentially affect HbA1c methodologies. This study compared the HbA1c measurements obtained in the presence and absence of hemoglobin variants using three platforms which have distinct analytical principles.
HbA1c (%) was measured using the Bio-Rad D-100 (high-performance liquid chromatography; reference method for this study), Abbott Alinity c (enzymatic assay) and Roche cobas c503 (immunoassay). Cohort 1 included 50 individuals with HbA1c values (3.5–5.9%) and no hemoglobin variants. Cohort 2 included 140 individuals with hemoglobin variants (3.5–10.4%), characterized by the Bio Rad D 100 or Sebia Capillarys systems. Variants included hemoglobinopathies/thalassemias (22 cases) and/or elevated HbF (5–30%). Agreement between methods was assessed using Deming regression and Bland–Altman plots with D-100 as the comparator. For samples with isolated HbF elevation (excluding structural variants), linear regression evaluated the effect of HbF (%) on HbA1c bias for Alinity and Roche. The percentage of reported or out-of-range/suppressed values for hemoglobin variants was compared.
In cohort 1, both Alinity and Roche showed strong agreement with D-100. Roche demonstrated average relative bias 1.11%; correlation coefficient r=0.938 while Alinity exhibited average relative bias 2.78%; r=0.938. In cohort 2, greater bias was observed with the presence of variants: Roche (bias -5.85%; r=0.954) and Alinity (bias -3.88%; r=0.960). For samples exhibiting only elevated HbF, both systems demonstrated a negative proportional effect relative to D-100 (Alinity: -0.052% HbA1c per 1% HbF; Roche: -0.057% per 1% HbF). Among variant samples with known hemoglobinopathies/thalassemias, D-100 suppressed results for 15.7% of cases (manually performed in lab) while Alinity did not report a specific value due to results being out of range in 12.9%, and Roche reported values for all samples.
The D-100, Alinity, and Roche platforms demonstrated high comparability in HbA1c samples without hemoglobin variants. In the presence of variants, particularly elevated HbF, both Alinity and Roche showed modest negative proportional bias relative to D-100 but maintained a strong correlation.
Comparing HbA1c results obtained by PEIT with those from HPLC to evaluate agreement and potential clinical interchangeability in patients with common haemoglobin variants shows acceptable agreement without evidence of substantial systematic or proportional error.
A. Mir, Koushik Biswas· Clinical Chemistry· 0 citations
Hemoglobin A1c (HbA1c) is a critical biomarker for diabetes diagnosis and monitoring long-term glycemic control. Separation-based methods, such as the Bio-Rad D-100 ion-exchange HPLC and the Sebia Capillarys® Octa 3 capillary electrophoresis system, provide excellent resolution of hemoglobin species and variants,...
Shu-Ling Fan, W. Gray· Clinical Chemistry· 0 citations
Hemoglobin A1c (HbA1c) represents average glycemia over ∼2–3 months based on glycation of HbA. Capillary electrophoresis (CE) separates Hb fractions based on charge and produces electropherograms that allow for integrated flags and interpretive comments when interferences or Hb variants are detected. ARUP Laborat...
Wen-Han Cheng, Kelly Doyle· Clinical Chemistry· 0 citations
HbA1c% remains the primary diabetes screening and long-term monitoring analyte as it mirrors the long-term damage done by protein glycation under chronically elevated glucose levels, as demonstrated by the landmark DCCT study. However, this testing assumes a patient with normal hemoglobin expression; predominantl...
Deanna D. H. Franke, M. Wagner· Clinical Chemistry· 0 citations
Hemoglobin A1c (HbA1c) is the primary marker for monitoring long term glycemic control, and harmonization across testing methods remains essential for clinical decision-making. Roche recently introduced the Tina quant® Hemoglobin A1cDx Gen. 4 (A1CX4) assay, a monoclonal antibody–based formulation that replaces th...
A. El Zubeir, D. Nyholt, Tony Chetty et al.· Clinical Chemistry· 0 citations
Background: Glycated haemoglobin (HbA1c) occupies an unchallenged position as the premier biomarker for the diagnosis and longitudinal monitoring of diabetes mellitus. Its clinical utility, however, depends entirely on the analytical fidelity of the measurement platform in use. Haemoglobin variants—encompassing both st...
Anush N, Sukhdeep Kumar, Jasmin Anush et al.· ICME JOURNAL OF LABORATORY M...· 0 citations