Video photoplethysmography (vPPG) uses a digital camera to detect pulsations in the blood vessels of the skin, which can be used to estimate various physiological parameters. The aim of this study was to investigate the accuracy and precision of heart rate variability (HRV) indices estimated using a smartphone camera and facial vPPG technology. The study group included 35 healthy volunteers (17 females) with median age 25 years (range 20–42 years). The subjects were in a sitting position, keeping their heads relatively still. A smartphone mounted on a tripod was used to acquire 1-min video recordings of participants’ faces. In parallel, a 1-lead chest electrocardiogram (ECG) was recorded to obtain reference values of two analysed time-domain HRV indices: SDNN and lnRMSSD. For SDNN, the median absolute error was 3.0 ms (7.1% in relative terms), and the root-mean-square error was 4.5 ms (15.7%). For lnRMSSD, the median absolute error was 0.20 (5.6%) and the root-mean-square error was 0.31 (9.9%). Correlations between the vPPG-based and ECG-based HRV values were strong, with the Pearson correlation coefficient of 0.98 for SDNN and 0.88 for lnRMSSD (
P
< 0.001 in both cases). In a young, white population, the tested vPPG technology estimated the HRV indices (SDNN and lnRMSSD) with good overall accuracy, with a slight systematic overestimation, especially for low values. The results should be confirmed in a larger study with greater diversity in age and skin tone.
In a young, white population, the tested vPPG technology estimated the HRV indices (SDNN and lnRMSSD) with good overall accuracy, with a slight systematic overestimation, especially for low values.
Leszek Pstras, T. Okupnik, B. Ponikowska et al.· Scientific Reports· 1 citation· ⚡1
Contactless vital-sign monitoring from ordinary cameras can expand access to physiological assessment in telehealth and consumer devices. A key limitation is variability in signal quality driven by illumination, motion, and skin phototype, which can reduce robustness and fairness.
To describe and evaluate...
A. Drohomirecka, P. Jaworski, S. Sobczak et al.· European Heart Journal· 0 citations
Wrist PPG reliability is condition dependent, and an interpretable signal-based reliability flag may complement point heart-rate estimates, and an interpretable signal-based reliability flag may complement point heart-rate estimates in previously unseen participants.
Angel Xu· Theoretical and Natural Scie...· 0 citations
The Muse 2 EEG headset also records forehead photoplethysmography (PPG). We tested whether its pulse intervals can replace Polar H10 RR intervals for ultra-short heart-rate variability in 28 seated recordings (60 s eyes open, then 60 s eyes closed), using fixed processing, no exclusions and eleven sensitivity analyses....
I. P. A. Griadhi, I. A. A. Suadnyana, N. Wahyuni· medRxiv· 0 citations
A real-time SpO2 estimation algorithm based on chest reflectance PPG that supports the technical feasibility of chest reflectance PPG-based SpO2 estimation and its accuracy under controlled desaturation conditions is developed and clinically evaluated.
Woo-Yong Lee, Jaeyeon Shin, Mih-Ye Song et al.· Italian National Conference...· 0 citations