Beyond Average Accuracy: Motion-Related Failure Patterns in Wrist-Worn Photoplethysmographic Heart-Rate Monitoring

Sep 2026 · Theoretical and Natural Science · 0 citations

TL;DR

Wrist PPG reliability is condition dependent, and an interpretable signal-based reliability flag may complement point heart-rate estimates, and an interpretable signal-based reliability flag may complement point heart-rate estimates in previously unseen participants.

Abstract

Wrist-worn photoplethysmography (PPG) enables unobtrusive heart rate monitoring, but motion-related errors tend to cluster under specific conditions and are often masked by aggregate accuracy metrics. This study characterized those failure patterns and assessed whether wrist-sensor features could identify unreliable estimates in previously unseen participants. PPG-DaLiA recordings from 15 participants, acquired with an Empatica E4 and synchronized chest electrocardiography during daily-life activities, were segmented into 8-s windows with a 2-s step. Heart rate was estimated using a transparent dominant-frequency PPG method. A large error was defined a priori as an absolute deviation of more than 10 beats per minute (bpm) relative to the ECG-derived reference. Motion intensity, acceleration-band power, PPG-acceleration spectral overlap, peak dominance and spectral entropy were used as reliability predictors. Of 47,182 non-transition windows, 14,622 (31.0%) met the large-error criterion. Failure prevalence differed substantially by activity, from 5.8% during sitting to 63.8% during stair climbing. In leave-one-subject-out validation, a motion-only model achieved an AUROC of 0.743, wrist-signal features increased AUROC to 0.851, and activity context increased it to 0.864. Wrist PPG reliability is therefore condition dependent, and an interpretable signal-based reliability flag may complement point heart-rate estimates. External prospective validation is required before clinical or consumer deployment.

Read PDF

Similar papers

Open access 2026

Interpretable Reliability Control for Wearable PPG Heart Rate Estimation

Wearable photoplethysmography (PPG) can return a plausible heart rate (HR) after motion or poor optical coupling has made the measurement unreliable. We treat HR reporting as a reliability-control problem and introduce an interpretable signal quality index (SQI) layer that reports coverage and residual error. On PPG-Da...

Ye-Kun Liang, Shi-Wei Zhang, Zhou Yang · 0 citations
Open access Sep 2026

Real-Time SpO2 Estimation from Chest Reflectance Photoplethysmography: Algorithm Development and Clinical Validation Against Arterial SaO2

A real-time SpO2 estimation algorithm based on chest reflectance PPG that supports the technical feasibility of chest reflectance PPG-based SpO2 estimation and its accuracy under controlled desaturation conditions is developed and clinically evaluated.

Woo-Yong Lee, Jaeyeon Shin, Mih-Ye Song et al. · 0 citations
Conference Open access 2026

Accuracy of Wearable Heart Rate Monitors under Static and High-Intensity Activity Conditions

The widespread adoption of wearable devices has created new opportunities for continuous monitoring of physical activity, cardiovascular health, and sleep patterns. However, it remains an open question to what extent the physiological and kinematic data provided by commercial smartwatches can be considered accurate and...

János Dávid Balogh, Attila Adamkó · 0 citations
Review Open access Oct 2026

Beyond Atrial Fibrillation: Emerging Clinical Applications of Photoplethysmography in Consumer Wearables

Introduction: Photoplethysmography (PPG) is an optical sensing technology embedded in many current consumer smartwatches and smart rings. Long used for heart-rate monitoring and pulse oximetry, PPG has more recently become the foundation for population-scale screening for atrial fibrillation. The clinical applications...

Muhammad Rifqi Haikal · 0 citations
Book Open access Oct 2026

Comparing Self-Supervised Pretraining Methods and Transfer Configurations for Wrist ECG under Motion

Wrist-worn electrocardiogram (ECG) enables unobtrusive and continuous cardiac monitoring in daily life, but reliable heart rate (HR) and heart rate variability (HRV) estimation remains challenging due to limited labeled data and electrode placement area, and severe motion artifacts. Self-supervised learning (SSL) provi...

Yan-Qiu Hu, Zhi-Yu Jia, Si-Ying Chen et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.