Real-Time SpO2 Estimation from Chest Reflectance Photoplethysmography: Algorithm Development and Clinical Validation Against Arterial SaO2

Sep 2026 · Italian National Conference on Sensors · Vol 26 · 0 citations · 66 references
Medicine

TL;DR

A real-time SpO2 estimation algorithm based on chest reflectance PPG that supports the technical feasibility of chest reflectance PPG-based SpO2 estimation and its accuracy under controlled desaturation conditions is developed and clinically evaluated.

Abstract

Peripheral oxygen saturation (SpO2) estimation using photoplethysmography (PPG) is typically based on the pulsatile components of red and infrared (IR) PPG signals acquired from peripheral sites, such as the finger or earlobe. Although these sites provide strong PPG signals, they are less suitable for integrated monitoring with electrocardiography (ECG), including arrhythmia detection. Simultaneous acquisition of PPG and ECG from a chest-worn device may therefore enable continuous and integrated physiological monitoring. However, chest PPG is characterized by lower perfusion and greater susceptibility to respiratory and motion artifacts, making reliable extraction of pulsatile red and IR components challenging. In this study, we developed and clinically evaluated a real-time SpO2 estimation algorithm based on chest reflectance PPG. Green PPG, which provides a relatively distinct pulsatile waveform under low-perfusion conditions, was used for beat detection. A heart-rate-adaptive analysis window, normalized cross-correlation-based signal quality assessment, and least-squares estimation of the AC amplitude ratio were combined to calculate the red-to-IR ratio of ratios, from which SpO2 was estimated using a calibration equation. Algorithm performance was evaluated against reference arterial oxygen saturation (SaO2) measured using a blood gas analyzer. The proposed method achieved an accuracy root mean square of 2.91% relative to the reference SaO2. These results support the technical feasibility of chest reflectance PPG-based SpO2 estimation and its accuracy under controlled desaturation conditions.

Read PDF

Similar papers

Preprint Aug 2026

CardioFusion-AI: Robust ECG--PPG Fusion for Multimodal Physiological Monitoring Under Signal Degradation

Wearable electrocardiogram (ECG) and photoplethysmogram (PPG) sensors are complementary but individually fragile: motion artifact, poor contact, and sensor dropout can degrade one or both signals. Fusion strategies that assume both modalities are equally trustworthy can become less reliable than a single clean modality...

Navaneetha Krishnan Kamalakannan, J. Kamalakannan · 1 citation
Open access Sep 2026

Beyond Average Accuracy: Motion-Related Failure Patterns in Wrist-Worn Photoplethysmographic Heart-Rate Monitoring

Wrist PPG reliability is condition dependent, and an interpretable signal-based reliability flag may complement point heart-rate estimates, and an interpretable signal-based reliability flag may complement point heart-rate estimates in previously unseen participants.

Angel Xu · 0 citations
Open access Sep 2026

Development and validation of an integrated wearable system for continuous cardiorespiratory monitoring using photoplethysmography and optical respiratory detection.

Cardiovascular diseases remain the leading cause of mortality worldwide, with ischaemic heart disease the leading contributor and acute myocardial infarction accounting for the majority of ischaemic heart disease deaths. This study presents the development and technical validation of a low-cost, integrated wearable sys...

Jesús David Ramírez, N. Toro, Juan Manuel Devaldenebro Campo et al. · 0 citations
Open access 2025

Evaluation of Heart Rate Variability by Smartphone App Using Pulse Photoplethysmography in Acute Myocardial Infarction

Highlights Smartphone-based PPG measurements of SDNN and rMSSD effectively differentiate AMI patients from healthy controls, with SDNN ≤21.35 ms showing high sensitivity (87.6%) for detection. Reduced HRV indices (SDNN and rMSSD) correlate with clinical severity markers, including lower LVEF, higher Killip class, and m...

Tien Anh Hoang, Nhat Long Tran · 1 citation
Open access Sep 2026

Validation study on cuffless ambulatory blood pressure monitoring method using bio-impedance, impedance cardiography and electrocardiography.

PURPOSE Continuous, non-invasive blood pressure (BP) monitoring is essential for understanding cardiovascular dynamics in real-world settings. This study evaluated the reliability and validity of a cuffless method for ambulatory BP monitoring using pulse wave travel times derived from bio-impedance, impedance cardiogra...

Myrte Schoenmakers, Martin J. Gevonden, B. Sauce et al. · 0 citations
Oct 2026

Comparative Assessment of PPG Signal Quality Across Fingers for Wearable Health Monitoring and Implications for Sensor Design Optimization

In clinical settings, every minute of delay in treating a cardiopulmonary emergency can reduce a patient's chance of survival by up to 10%. Photoplethysmography (PPG) is a low-cost, noninvasive method widely adopted in both clinical, and consumer settings to monitor cardiopulmonary signals. Therefore, clinical PPG puls...

Karlen Aleksanyan, Xue Xu, Zheng Liu et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.