Optimizing IVIM measurement on magnetic resonance linear accelerator using multi-b-value acquisition and physics-informed neural network.

Sep 2026 · Physica medica (Testo stampato) · Vol 150, pp. 107205 · 0 citations · 35 references
Medicine

TL;DR

The combination of multi-b-value acquisition and PINN provided a robust and accurate framework for IVIM parameter estimation on the MR-Linac and showed potential for integration into MR-Linac quantitative imaging workflows.

Abstract

Background

Accurate quantification of intravoxel incoherent motion (IVIM) parameters on the magnetic resonance linear accelerator (MR-Linac) is challenging due to low signal-to-noise ratio and time constraints.

Purpose

This study aimed to evaluate the impact of b-value acquisition strategies and estimation algorithms on IVIM parameter measurement to guide reliable IVIM quantification on the MR-Linac.

Methods

A digital phantom with high number of signals averaged (high-NSA) and multi-b-value signals was constructed to evaluate estimation accuracy using root mean square error (RMSE). Ten healthy volunteers underwent four repeated diffusion-weighted imaging (DWI) scans, consisting of two high-NSA DWIs and two multi-b-value DWIs to assess repeatability via the within-subject coefficient of variation (wCV). IVIM parameters from high-NSA DWI were estimated using Segment fitting, while multi-b-value DWI data were processed using least squares (LS), Bayesian estimation, and physics-informed neural network (PINN).

Results

All IVIM parameters from multi-b-value DWI exhibited lower wCVs than those from high-NSA DWI, except for the LS-derived diffusion coefficient (D). Multi-b-value DWI combined with PINN achieved rapid and highly repeatable estimation. In the digital phantom, high-NSA DWI with Segment fitting yielded the highest RMSE for the pseudo-diffusion coefficient (D∗). LS and Bayesian methods resulted in higher D RMSEs and required longer fitting times. In contrast, multi-b-value DWI + PINN achieved the lowest RMSEs across all parameters with minimal computational cost.

Conclusion

The combination of multi-b-value acquisition and PINN provided a robust and accurate framework for IVIM parameter estimation on the MR-Linac and showed potential for integration into MR-Linac quantitative imaging workflows.

Read PDF

Similar papers

Open access Sep 2026

Technical Feasibility of IVIM for Prostate Cancer Imaging Using High‐Amplitude Whole‐Body Gradient System With Spiral Readout and Velocity‐Compensated Diffusion Encoding

Intravoxel incoherent motion (IVIM) is sensitive to tissue diffusion and perfusion and has spurred great interest in various pathologies such as prostate cancer. However, poor repeatability and reproducibility of IVIM parameters hamper its wide‐scale adoption in the clinics. To address this, improvements to the IVIM da...

Malwina Molendowska, Ivan A. Rashid, K. Foley et al. · 0 citations
Open access Sep 2026

Evaluations of retrospective frequency and phase correction methods for single-voxel MR spectroscopy at 7T

Subject motion and gradient heating-induced frequency and phase offsets result in spectral misalignment during single-voxel MR spectroscopy (SVS) acquisitions. Several methods have been presented to align the spectra, but their performance on a 7T system is unclear. This study aimed to evaluate the practical importance...

Chu-Yu Lee, Jia Xu, Bao-Lian Yang et al. · 0 citations
Open access Sep 2026

High Dynamic Range Turbo-FLASH B1 Mapping for Characterizing Multi-Transmit Multi-Receive RF Arrays, Validated at 7 T and 11.7 T.

This work presents a new B1 + and B1 - model-based fitting approach for interferometric presaturated TurboFLASH data, which improves B1 + and B1 - mapping robustness to noise and enables more flexible satTFL acquisition scheme, allowing better trade-offs between accuracy and scan time.

Natalia Dudysheva, Franck Mauconduit, Michel Luong et al. · 0 citations
Sep 2026

Benchmarking 18F and 68Ga Phantom Performance on a Digital Time-of-Flight PET/CT System: Reference Values for Spatial Resolution, Contrast Recovery, and Background Variability.

Under matched reconstruction conditions, 18F retained superior intrinsic resolution and contrast recovery, whereas longer acquisitions mainly reduced noise and could not overcome the physical resolution limitations of 68Ga.

Amer H. Ghujeh, M. Altinyay, Antar Aly et al. · 0 citations
Sep 2026

High-dose-rate prostate brachytherapy imaging optimization: A motorized sagittal three-dimensional ultrasound acquisition system.

The motorized attachment for SR3D acquisition reduced image variability and improved image quality, and may encourage broader use of sagittal reconstruction in HDR-BT, potentially improving patient outcomes.

David Contella, Chandima Edirisinghe, Kevin Barker et al. · 0 citations
Open access Oct 2026

Qualitative and Quantitative Comparison of Readout-Segmented and Single-Shot Echo-Planar Diffusion-Weighted MR Imaging of the Prostate At 3.0 T

Abstract Aims and objectives The aim of the study is to qualitatively and quantitatively compare the diagnostic value of diffusion-weighted (DW) magnetic resonance (MR) imaging based on standard single-shot echo-planar imaging (ss-EPI) and readout-segmented echo-planar imaging in patients with prostate cancer at 3.0 Te...

Thiagarajan Veerappan, B. P. Sathyanathan · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.