Benchmarking 18F and 68Ga Phantom Performance on a Digital Time-of-Flight PET/CT System: Reference Values for Spatial Resolution, Contrast Recovery, and Background Variability.

Sep 2026 · Journal of Nuclear Medicine Technology · pp. jnmt.126.272820 · 0 citations · 22 references
Medicine

TL;DR

Under matched reconstruction conditions, 18F retained superior intrinsic resolution and contrast recovery, whereas longer acquisitions mainly reduced noise and could not overcome the physical resolution limitations of 68Ga.

Abstract

Our purpose was to establish reference performance benchmarks for spatial resolution, contrast recovery coefficient (CRC), and background variability (BV) for 68Ga and 18F under matched conditions and to determine the extent to which acquisition duration can compensate for isotope-specific resolution differences. Methods: Phantom studies were performed on a digital PET/CT platform using capillary sources in air (minimal scatter) and a triple-line insert in water (scatter conditions). A National Electrical Manufacturers Association International Electrotechnical Commission body phantom (spheres 10-37 mm; 4:1 activity ratio) was used to assess CRC (International Atomic Energy Agency definition) and BV (coefficient of variation). Data were analyzed using Kruskal-Wallis tests for resolution and 2-way ANOVA for CRC/BV, evaluating the interaction between isotope and acquisition duration. Results: 18F consistently yielded resolution limits (full width at half maximum) superior to those of 68Ga in both air and water (Kruskal-Wallis, P ∼ 0.008). Notably, extending the acquisition duration from 2 to 5 min provided no significant improvement in full width at half maximum for either tracer (P = 0.70). In National Electrical Manufacturers Association testing, 18F maintained a significant CRC advantage across all sphere sizes and durations. At 10 min, the CRC for the 37-mm sphere reached about 97.5% for 18F versus about 81.3% for 68Ga. For small spheres (10-13 mm), 18F exceeded 68Ga recovery by about 14%-22%. Although increasing scan duration from 2 to 10 min significantly reduced BV (noise) for both tracers (18F: 1.92%→1.37%; 68Ga: 2.47%→1.29%; P = 0.026), it did not bridge the CRC gap. Conclusion: This work provides practical benchmark values for 18F and 68Ga on a digital PET/CT platform. Under matched reconstruction conditions, 18F retained superior intrinsic resolution and contrast recovery, whereas longer acquisitions mainly reduced noise and could not overcome the physical resolution limitations of 68Ga. These data support quality assurance, protocol optimization, and harmonization of prostate-specific membrane antigen PET imaging across both tracers.

View source

Similar papers

Sep 2026

EARL-based quantitative evaluation of administered activity reduction for [68Ga]Ga-DOTATOC PET/CT on a 32-cm axial FOV digital BGO system: Phantom analysis and retrospective list-mode clinical assessment.

PURPOSE The increased sensitivity of long-axial field-of-view (FOV) digital PET/CT systems enables potential reductions in administered activity and/or acquisition time with respect to standard-field-of-view systems and established clinical protocols, while preserving quantitative performance. This study aimed to evalu...

A. Piai, B. Bordigoni, L. Antunovic et al. · 0 citations
Open access Sep 2026

Effects of count rate performance on image quality in an LAFOV PET/CT system

This study investigated the relationship between body size, administered activity, and image quality (IQ) using two patient-like phantoms representing normal and obese body profiles, and demonstrated the count rate capabilities up to and beyond the saturation limit.

Nanna Overbeck, Zuzanna Czeluśniak, Søren Holm et al. · 0 citations
Sep 2026

Characterization of open-source BSREM reconstruction for 123I-FP-CIT spect on a conventional anger-type SPECT/CT system: a phantom study.

Open-source BSREM showed task-, region, and count-dependent behavior, supporting task- and count-specific parameter selection before clinical application, and showed reduced left-putaminal uptake remained underestimated, and putaminal asymmetry moved further from their theoretical values.

Takuro Shiiba, Shingo Tanahashi, Kohei Takasu et al. · 0 citations
Oct 2026

Multi-center comparison of relative stopping power estimation of an ex vivo phantom using clinical X-ray CT scanners and a proton CT prototype.

BACKGROUND AND PURPOSE Accurate estimation of relative stopping power (RSP) is essential for proton therapy (PT) treatment planning. Emerging X-ray computed tomography (xCT) technologies are enabling more advanced RSP calibration strategies, but may also introduce greater variability across centers. This study aimed to...

E. Fogazzi, E. Bär, M. Bruzzi et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.