The Infrascanner showed good sensitivity but modest specificity for ICH detection in this two-center Pakistani cohort, consistent with published series.
Abstract
Background
AND
Objectives
Intracranial hemorrhage (ICH) is a major cause of morbidity and mortality after traumatic brain injury (TBI). Computed tomography (CT) is often unavailable in low-resource settings, delaying diagnosis. The Infrascanner 2000™, a handheld near-infrared device, has been proposed as a rapid, noninvasive ICH screening tool. The objective of this study was to determine the diagnostic accuracy of the Infrascanner 2000™ for detecting ICH in TBI patients, compared with noncontrast CT, at 2 tertiary care hospitals in Peshawar, Pakistan.
Methods
In this prospective observational study, 141 patients with mild, moderate, or severe TBI presenting to the emergency department of Neurosurgery at Northwest General Hospital and Lady Reading Hospital, Peshawar, Pakistan, underwent Infrascanner examination (inter-rater kappa 0.82), followed by noncontrast CT as the reference standard. Sensitivity, specificity, predictive values, likelihood ratios, balanced accuracy, and overall accuracy were calculated. Balanced accuracy (mean of sensitivity and specificity) replaced area under the curve, as the device yields a single binary operating point. Results followed Standard for Reporting Diagnostic Accuracy Studies guidelines.
Results
CT confirmed hemorrhage in 125 of 141 patients (88.6%). The Infrascanner showed sensitivity of 91.2%, specificity of 50.0%, positive predictive value of 93.4%, negative predictive value of 42.1%, and overall accuracy of 86.5% (positive likelihood ratio 1.82; negative likelihood ratio 0.18). Balanced accuracy was 0.706 (95% CI: 0.626-0.786), a single operating point rather than a conventional receiver operating characteristic-derived area under the curve. True positives were predominantly superficial hematomas; false negatives may reflect small or deep lesions. False positives could reflect extracranial soft-tissue injury, though this was not formally graded.
Conclusion
The Infrascanner showed good sensitivity but modest specificity for ICH detection in this two-center Pakistani cohort, consistent with published series. It may serve as a rapid triage adjunct where CT access is delayed but should not replace CT imaging. Further multicenter studies, including prehospital settings, are warranted.
Background: Mild traumatic brain injury (mTBI) accounts for the majority of traumatic brain injury presentations in emergency departments, yet its diagnostic assessment remains challenging because clinically significant intracranial lesions are relatively uncommon, while missed injuries may still lead to important clin...
Vitor Gabriel Rodrigues de Oliveira, M. L. R. Bastos da Silva· ANPAbstract_CBN26· 0 citations
Objective: To compare the diagnostic accuracy of InfraScanner against computed tomography (CT) for detection of intracranial haematoma in paediatric versus adult patients with traumatic brain injury (TBI), and to identify lesion-level predictors of device detection.
Methodology: This prospective observational, CT-vali...
Muhammad Nawaz Khan, Syed Shayan Shah, Muhammad Sohaib Khan· Pakistan Journal of Medical...· 0 citations
Background.
Severe traumatic brain injury (TBI) remains one of the leading causes of mortality and disability, particularly in resource-limited settings where invasive intracranial pressure (ICP) monitoring is unavailable. The international Brain Trauma Foundation guidelines require specialized equipment, creating a...
Background: Severe traumatic brain injury (sTBI) remains a major cause of mortality and disability in children worldwide. Intracranial pressure monitoring (ICPm) and decompressive craniectomy (DC) are commonly employed in the management of pediatric sTBI, the clinical and radiological criteria guiding these interventio...
Angelo R. Silva, Derick Pedrosa Pachá, Rui Manuel Morais de Deus et al.· Archives of Pediatric Neuros...· 0 citations
Traumatic brain injury (TBI) is the leading cause of death and disability in people under 40 worldwide. There is an unmet need for detection and treatment of life-threatening TBI when computed tomography (CT) scanners and neurosurgeons are unavailable. We developed a new handheld frequency-domain multidetector infrared...
D. Brody, Bradley A. Dengler, Joseph P. Angelo et al.· bioRxiv· 0 citations
We use cookies to run the site and, with your consent, for analytics and to show ads.
See our Cookie Policy.