Aug 2026· JIM - Journal of Immunological Methods· pp.
114112
· 0 citations· 17 references
Medicine
TL;DR
These locally derived RIs may improve the interpretation of FXII and ATIII results and reduce unnecessary follow-up testing caused by inappropriate generic intervals and highlight the importance of local RI verification when applying manufacturer-derived intervals or changing analytical platforms.
Abstract
Objective
This study aimed to establish and verify reference intervals (RIs) for coagulation factor XII (FXII) and antithrombin III (ATIII) in an adult population from the Jiangsu region of China and to evaluate the influence of analytical platform on RI interpretation.
Methods
We retrospectively analyzed data from 301 apparently healthy adults attending the Health Examination Center of Nanjing Drum Tower Hospital. Plasma FXII and ATIII activities were measured in parallel using two automated coagulation analyzers, the Sysmex CS-5100 and CS-6500. Statistical analysis followed CLSI EP28-A3 guidelines. Distribution normality was assessed to determine the appropriate method for RI establishment: the non-parametric percentile method for non-normally distributed data and the parametric method for normally distributed data. Analyzer-related agreement was further assessed using Bland-Altman analysis. The derived RIs were verified using an independent sample set of 20 apparently healthy adults.
Results
FXII showed a non-Gaussian distribution, whereas ATIII showed an approximately Gaussian distribution. For the CS-5100 analyzer, the RI was 32.2%-109.1% for FXII and 84.1%-104.7% for ATIII. For the CS-6500 analyzer, the corresponding RIs were 37.1%-111.1% for FXII and 86.1%-106.7% for ATIII. The newly established RIs differed from the manufacturer-provided ranges currently used in routine practice. No clinically meaningful sex-based partitioning was identified. Although analyzer-related differences were statistically significant, Bland-Altman analysis showed modest mean biases between the CS-6500 and CS-5100 platforms, with mean differences of +5.09% for FXII and + 1.17% for ATIII. All verification samples fell within the proposed RIs.
Conclusion
This study established and verified regional RIs for FXII and ATIII in healthy adults from Jiangsu, China. Although analyzer-related differences between the CS-5100 and CS-6500 platforms were modest, the findings highlight the importance of local RI verification when applying manufacturer-derived intervals or changing analytical platforms. These locally derived RIs may improve the interpretation of FXII and ATIII results and reduce unnecessary follow-up testing caused by inappropriate generic intervals.
The two analyzer–reagent systems showed positive analytical associations, but the magnitude and dispersion of paired differences varied by assay, but the magnitude and dispersion of paired differences varied by assay.
Betül Özbek Kurtul, Bağnu Dündar, G. Gündoğan et al.· Diagnostics· 0 citations
BACKGROUND
Growing interest surrounds circulating biomarkers for assessing neuromuscular integrity and muscle health. C-terminal agrin fragment (CAF) has emerged as a promising marker associated with sarcopenia, ageing, and renal dysfunction. Limited reference data and insufficient characterisation of preanalytical fac...
J. R. Therkildsen, C. S. Knudsen, J. Storgaard et al.· Clinica chimica acta; intern...· 0 citations
Interpretation of pediatric coagulation results depends on the reference interval (RI) applied, yet the extent to which RI source affects result classification is rarely quantified. This retrospective single-center study determined how RI source affects out-of-RI flagging in pediatric coagulation testing. Analyzer-spec...
Tuba Çakmak Ercan, D. Topcu, G. Bozkaya· Clinica chimica acta; intern...· 0 citations
INTRODUCTION
It is recommended to detect lupus anticoagulant (LA) using two phospholipid-dependent clotting assays, including a LA-sensitive activated partial-thromboplastin-time (aPTT). Among these reagents, Staclot-LA assay is performed on a mixture of test and normal plasma with and without hexagonal phase phospholi...
V. Siguret, G. Jourdi, S. Melicine et al.· International Journal of Lab...· 0 citations
Accurate hematological analysis is critical for clinical decision-making across diverse patient populations. This study evaluates the performance of the Yumizen H550, a compact hematology analyzer, by comparing it to the established Yumizen H2500 predicate device. The evaluation aimed to validate the H550’s accur...
Amin Mohammad, Azar Khatibi, J. Hipólito et al.· Clinical Chemistry· 0 citations
The study objectives were: 1) establish 99th percentile upper reference limits (URL) for the VITROS™ Immunodiagnostic Products hs Troponin I (VITROS hsTnI) Assay, 2) confirm VITROS hsTnI Assay meets analytical performance criteria for a high-sensitivity cardiac troponin assay established by the IFCC Task Force on...
A. Ott-Vasconi, Michael Colvin, S. Jackson et al.· Clinical Chemistry· 0 citations
We use cookies to run the site and, with your consent, for analytics and to show ads.
See our Cookie Policy.