Photon-counting detector CT with virtual monoenergetic imaging and iterative metal artifact reduction for spinal fusion implants: optimal keV and combined benefits.
Sep 2026· European Journal of Radiology· Vol 205, pp.
113250
· 0 citations· 27 references
Medicine
TL;DR
For spinal implants, PCD-CT offers effective approaches for metal artifact and 120 keV is preferred and the combination of iMAR and 120 keV introduce new artifacts and requires careful consideration.
Abstract
Objectives
To evaluate the effectiveness of virtual monoenergetic imaging (VMI) and iterative metal artifact reduction (iMAR), alone and in combination, for reducing metal artifacts in postoperative CT scans of patients with spinal fusion, using a clinical photon-counting detector CT (PCD-CT) system.
Materials And Methods
Consecutive patients after spinal instrumentation were prospectively included in this study and received PCD-CT imaging of spine. Image reconstructions comprised conventional polychromatic images (PI), virtual monoenergetic images (VMI) at 100-140 keV, and iterative metal artifact reduction (iMAR) for determining the optimal keV. Further a combined optimal keV VMI with iMAR was reconstructed. Objective image quality was quantified by image noise and artifact index (AIx) measured across five predefined anatomical regions. Subjective assessment was performed independently by two radiologists using a 5-point Likert scale to grade artifact severity and the visibility of adjacent anatomical structures.
Results
66 patients (32 males, 34 females; 60.5 ± 14.2 years) with a total of 70 scans were enrolled. All VMI reconstruction provided lower noise, lower AIx and better visibility and diagnostic confidence for both vertebra and adjacent anatomical structures than polychromatic images (P < 0.0001). Subjective and objective assessments in first step showed that 120 keV provided superior overall imaging quality. The further combination of 120-keV VMI with iMAR resulted in significantly lower metal artifacts but introduced new artifacts.
Conclusions
For spinal implants, PCD-CT offers effective approaches for metal artifact and 120 keV is preferred. The combination of iMAR and 120 keV introduce new artifacts and requires careful consideration.
In patients with bladder cancer undergoing PCD CT, diagnostic performance for muscle invasion was greater for 40-keV VMIs than for conventional polychromatic images.
Wei Wang, Jiangkai He, Jing-Yun Wu et al.· AJR. American journal of roe...· 0 citations
The findings support implementing standardized protocols with automatic dual-range reconstructions to enable efficient, individualized carotid plaque assessment in clinical practice and support implementing standardized protocols with automatic dual-range reconstructions for keV selection based on plaque composition.
Mirela Kostadinova, Lea B. Uebelacker, Tommaso D'Angelo et al.· Diagnostics· 0 citations
Although DE-CBCT with VMI and frequency splitting improves HU agreement with CT, slightly improves soft tissue visualization, and reduces artifacts in head and neck phantom imaging, it did not reveal any additional targets compared with routine CBCT.
D. Jayakody, V. Riekki, M. Bode et al.· Biomedical engineering and p...· 0 citations
PURPOSE
To compare image-quality and radiation dose between photon-counting CT (PCD CT) and cone-beam CT (CBCT) for postoperative cochlear implantation (CI) imaging.
METHODS AND MATERIALS
This retrospective study included 224 postoperative CI datasets from July 2019 to February 2026, including 112 PCD CT scans using...
Iris Burck, M. Tillmanns, S. Bernatz et al.· European Journal of Radiolog...· 0 citations
Background/Objectives: This study aimed to investigate the impact of low-energy virtual monoenergetic imaging (VMI) on quantitative and qualitative image characteristics, as well as its effect on the diagnostic assessability of thoracic arteries in photon-counting computed tomography angiography (CTA). Methods: We retr...
Andreea I. Nica, C. Booz, G. Bucolo et al.· Diagnostics· 0 citations