A Tale of Two Calibrators: Comparing Newtonian and Photon Calibration in Advanced LIGO after the Third Observing Run

Sep 2026 · 0 citations · 11 references
Physics

Abstract

Precise calibration of LIGO's strain readout plays a vital role in our ability to extract information from gravitational-wave detections. We present a comparative analysis of two independent direct-force calibration methods: the Newtonian Calibrator (NCal) and Photon Calibrator (PCal) systems. Currently, LIGO relies solely on the PCal system as its absolute calibration reference. To facilitate comparison, a series of direct-force injections were performed using the NCal and PCal systems at the LIGO Hanford observatory just after the completion of the third observing run. We compute the ratio of the measured to expected strain amplitudes for every injection. For each injection frequency across a 30 Hz band, the ratios of NCal and PCal differ by $\sim$0.5-1%, demonstrating a small systematic difference between the two systems. This offset highlights the importance of maintaining multiple independent calibration references to validate LIGO's calibration.