Validation of the Arm-Type Fully Automatic Blood Pressure Monitor DBP-62F4B-P in adolescents and adults according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (ISO 81060-2:2018+Amd 2:2024) standard.
Aug 2026· Blood Pressure Monitoring· 0 citations· 14 references
Medicine
TL;DR
The Arm-Type Fully Automatic Blood Pressure Monitor (model: DBP-62F4B-P) meets the accuracy and consistency requirements of the AAMI/ESH/ISO standard for SBP and DBP measurement in adolescents and adults with arm circumference 22.0-42.0 cm and is suitable for home use based on its design features and validation performance.
Abstract
Objective
To validate the accuracy and safety of the Arm-Type Fully Automatic Blood Pressure Monitor (model: DBP-62F4B-P) manufactured by JOYTECH Healthcare Co., Ltd. for blood pressure (BP) measurement in adolescents and adults with arm circumference 22.0-42.0 cm, in accordance with the Association for the Advancement of Medical Instrumentation (AAMI)/European Society of Hypertension (ESH)/International Organization for Standardization (ISO) (ISO 81060-2:2018+Amd 2:2024) international standard.
Methods
A total of 90 subjects were enrolled. Three valid datasets were collected from each subject, yielding 270 datasets for analysis. The mercury sphygmomanometer (reference device) data were used as the gold standard to evaluate whether the accuracy of the test device met the requirements.
Results
Based on 270 valid datasets, the mean difference between the test device and the reference device for SBP was -0.49 mmHg (SD = 7.29 mmHg), and for DBP was 1.45 mmHg (SD = 6.14 mmHg), meeting Criterion 1 requirements. Analysis based on individual averages of the 90 subjects showed a mean difference of -0.49 mmHg (SD = 5.57 mmHg) for SBP and 1.45 mmHg (SD = 5.05 mmHg) for DBP, both meeting the requirements of Criterion 2. The device maintained stable measurement performance across different arm circumferences, ages, and BP levels.
Conclusion
The Arm-Type Fully Automatic Blood Pressure Monitor (model: DBP-62F4B-P) meets the accuracy and consistency requirements of the AAMI/ESH/ISO (ISO 81060-2:2018+Amd 2:2024) standard for SBP and DBP measurement in adolescents and adults with arm circumference 22.0-42.0 cm. It can be safely and effectively used for BP monitoring in clinical settings and, by extension, is suitable for home use based on its design features and validation performance.
The iHealth Track KN-550BT upper-arm oscillometric upper-arm automated BP monitor fully complies with the requirements of ISO 81060-2:2018+AMD1:2020/AMD2:2024, and can be recommended for clinical application and home self-BP measurement among pregnant patients.
Ying Zhang, Dongmei Wang, Chunmei Cui et al.· Blood Pressure Monitoring· 0 citations
This represents the first evaluation of a Device Type 3 cuffless blood pressure algorithm (automated, wearable, demographic-calibrated, not at heart level) meeting ESH awake/asleep test criteria, suggesting Alysis-001's potential as a clinically viable alternative for ambulatory blood pressure monitoring in hypertensiv...
K. Hongyo, Atsushi Hirayama, Kosuke Shimizu et al.· Hypertension Research· 0 citations
It is demonstrated that a training-phase evaluation of Type T mode continuous automated non-invasive sphygmomanometer repeated SBP, DBP, and MAP measurement of oCare™ BP100 in Type T mode are stable across various analysis periods and patients.
Chao-Lun Lai, Yeh-Wen Lee, Chia-Fang Chang et al.· Frontiers in Medical Technol...· 0 citations
The metabolic syndrome (MetS) is characterized by having three or more of these risk factors: increased blood glucose (Glc), serum triglycerides (TG), blood pressure (BP), waist circumference (WC), and reduced high-density lipoprotein (HDL). Mobile diagnostics using point-of-care (POC) blood analyzers could help identi...
S. Houtenbos, A. Schraplau, P. Wippert et al.· BMC Primary Care· 0 citations
The 2015–2025 ecosystem shows rapid progress and a transition toward cuffless-specific validation, however, clinical utility depends on intended use, population, calibration/recalibration, and setting, and performance should not be extrapolated across technologies, populations, or contexts.
J. C. Bustamante-Rodríguez, Jhosmer Ballena-Caicedo, Julio César Bautista-Zuta et al.· Frontiers in Digital Health· 0 citations
The Microlife WatchBP O3 upper-arm, oscillometric BP device passed the criteria of the ISO 81060-2:2018 standard in normotensive pregnant women, women with gestational hypertension, and women with pre-eclampsia.
A. de Greeff, Catherine A Cluver, A. Shennan· Blood Pressure Monitoring· 0 citations
We use cookies to run the site and, with your consent, for analytics and to show ads.
See our Cookie Policy.