Detecting the Critical View of Safety in Laparoscopic Cholecystectomy with Deep-CNN
Abstract
Laparoscopic cholecystectomy ranks among the most commonly performed operations worldwide, yet bile duct injury remains a serious complication with potentially life-altering consequences for patients. Confirming the Critical View of Safety (CVS) before any clip is placed is the accepted safeguard against such injuries, but the assessment continues to rely heavily on individual surgical judgment. This paper describes and compares three convolutional network architectures-ConvNeXt-Tiny, EfficientNet-B3, and ResNet-50-trained on the Endoscapes2023 dataset to classify individual laparoscopic video frames as CVS achieved or not achieved. All three networks were trained under a two-phase transfer learning scheme that preserves pre-trained feature representations while steering the network toward surgical anatomy. Among the three networks tested, ConvNeXt-Tiny recorded the highest area under the receiver operating characteristic curve (AUC) of 0.7655 along with a recall of 0.65 on the positive class. This recall was obtained after calibrating the decision threshold to favour sensitivity, and while it makes ConvNeXt-Tiny the most defensible of the three networks for this task, a 35% miss rate on true CVS-achieved frames means the model should be regarded as a promising baseline rather than a stand-alone safety net; missing a safety landmark carries real clinical risk, and further validation would be required before any intraoperative use. A browser-based inference dashboard with live camera support and Gradient-weighted Class Activation Mapping (Grad-CAM) overlays was also built to make predictions interpretable and practically accessible.