Skip to content

A-367 Assessment of Hemolysis Interference on HbA1c Measurement Using the Siemens Healthineers DCA Vantage Analyzer

Oct 2026 · Clinical Chemistry · 0 citations

Abstract

The Siemens Healthineers DCA Vantage Analyzer is a widely used point of care analyzer for hemoglobin A1c (HbA1c) testing and employs whole blood specimens for routine measurement. Although whole blood assays inherently contain intracellular hemoglobin, the potential impact of pre-analytical in vitro hemolysis on DCA Vantage Analyzer HbA1c results has not been systematically evaluated, and no manufacturer defined acceptance criteria currently exist for hemolysis interference. This study assessed the analytical performance of the DCA Vantage Analyzer HbA1c assay in the presence of controlled degrees of hemolysis and evaluated whether hemolysis introduces clinically significant bias, given that hemolysis is a common pre-analytical issue in fingerstick samples. Fresh EDTA whole blood samples were pooled to create three HbA1c concentration groups (5.0–5.5%, 6.0–7.0%, and 10.0–12.0%). Red blood cells were separated from plasma to prepare hemolysate using a freeze–thaw method and quantitatively verified. A series of samples was generated ranging from 0% to 100% hemolysis for each HbA1c concentration group. All samples were analyzed on the DCA Vantage Analyzer (16 replicates per sample) and compared to the BioRad D100 HPLC method (3 replicates per sample). Deming and least squares regression were used to assess the relationship between hemolysis and HbA1c bias. Interference exceeding ±7% from the reference value was considered unacceptable per Siemens Healthineers specifications. Across all three HbA1c concentration pools, HbA1c measurements on the DCA Vantage Analyzer remained stable throughout the entire hemolysis range (0–100%). In the 5–5.5% and 6–7% pools, regression slopes were nearly flat, with changes of less than 0.2% HbA1c between non hemolyzed and fully hemolyzed samples. The 10–12% pool showed a slightly greater but still small positive trend (<0.3% absolute change in HbA1c) across hemolysis levels. At every hemolysis level, mean HbA1c values remained within the predefined ±7% acceptance region relative to the non hemolyzed baseline. Replicate scatter was consistent across hemolysis levels, indicating that analytical precision of the DCA Vantage Analyzer was not affected by hemolysis. Collectively, these results demonstrate that the DCA Vantage Analyzer HbA1c assay is highly resilient to hemolysis, with no clinically meaningful interference even at extreme hemolysis levels (Figure 1). This study provides a comprehensive assessment of hemolysis interference on HbA1c measurements using the Siemens Healthineers DCA Vantage Analyzer. Although HbA1c is inherently a whole blood assay, where hemolysis is often not visible or routinely assessed, this study confirms that even substantial in vitro hemolysis does not affect HbA1c results. These findings eliminate concern regarding hemolysis related bias in DCA Vantage Analyzer HbA1c testing and support reliable interpretation of point of care results in the presence of hemolysis.

View source

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.