Nov 2026· Frontiers of Physics· Vol 14· 0 citations· 52 references
Abstract
Photoacoustic spectroscopy enables compact and selective gas sensing, but its weak acoustic signals require effective phase-sensitive detection. Conventional PAS systems commonly rely on standalone lock-in amplifiers, which separate the demodulation function from excitation generation, sensor control, and data management, limiting system integration and reconfigurability.
This work presents a real-time software-defined digital lock-in amplifier that unifies modulation-waveform generation, phase-coherent signal acquisition, quadrature demodulation, configurable filtering, real-time visualization, and data logging within a single LabVIEW application interfaced with a National Instruments USB-6361 data-acquisition module. The S-DLIA was integrated into a CO
2
PAS platform employing a distributed-feedback diode laser at 4992.5157 cm
-1
, a differential resonant photoacoustic cell, and second-harmonic wavelength modulation. It was benchmarked against a commercial SR830 lock-in amplifier processing the same photoacoustic signal under matched conditions.
The sensor responded linearly to CO
2
concentrations from 100 to 1,000 ppm (
R
2
= 0.999). Allan deviation analysis at an integration time of 220 s yielded minimum detection limits of 0.505 ppm for the S-DLIA and 8.909 ppm for the SR830 under the tested configuration, corresponding to a 17.6-fold lower detection limit compared with the SR830. Continuous operation over 24 h at a fixed laboratory location tracked indoor CO
2
variations without apparent drift.
These results validate the S-DLIA as a reconfigurable and flexible alternative to dedicated lock-in instruments in PAS, demonstrating the potential of software-defined synchronous detection for integrated gas-sensing systems.
A fiber‐coupled dual‐cell photoacoustic spectroscopy (PAS) sensing system for remote gas detection is proposed, in which the gas absorption region is physically separated from the acoustic detection region. A multi‐pass absorption cell with an effective optical path length of approximately 3 m serves as the absorption...
Chen Cao, Lei Li, Si-Yu Ding et al.· Microwave and optical techno...· 0 citations
We present a continuous-wave time-of-flight (CW-ToF) sensing system capable of megahertz-rate depth measurements using a free-space laser and commodity radio-frequency (RF) communication hardware. Existing highspeed optical displacement and vibration sensing systems typically rely on optical interferometry or specializ...
R. Bhaskara, Dhawal Sirikonda, Omkar Vengurlekar et al.· ACM Transactions on Graphics· 0 citations
Tunable diode laser absorption spectroscopy (TDLAS) has been widely applied in industrial gas detection, but traditional calibration methods face challenges including high standard gas consumption, exorbitant costs, and heavy calibration workloads—particularly under wide temperature and pressure ranges, where variation...
Tao Liu, Zhao-An Yu, Kai-Ping Yuan et al.· Photonics· 0 citations
To reduce the system volume, lower the instrument cost, and achieve highly sensitive multi‐gas measurement, a photoacoustic spectroscopic (PAS) gas detection system based on a non‐resonant photoacoustic (PA) cell was developed. The filter wheel, chopper and non‐resonant PA cell were designed. By comparing the feedback...
Jing-Hao Yang, Gong-Sheng Song, Chen Huang et al.· Microwave and optical techno...· 0 citations
Atmospheric methane (CH4) monitoring requires an optimal balance between ultra-high sensitivity and mechanical field-readiness. Conventional multi-pass architectures, such as Herriott cells, are fundamentally constrained by rigid geometries and alignment sensitivities that limit real-time optimization capability. To ov...