Skip to content
Open access

3-GHz Four-Channel Shift-Register Board for High-Rate, High-Precision Single-Photon Detection

2026 · IEEE Transactions on Instrumentation and Measurement · Vol 75, pp. 2005804-2005804 · 0 citations · 16 references

Abstract

Time-correlated single-photon counting (TCSPC) is the method of choice for high-resolution time-resolved measurements due to its exceptional temporal precision. Recent advances have demonstrated that TCSPC is no longer confined to the single-photon regime, as pile-up distortion, particularly severe under high illumination, can be accurately corrected by treating detector dead time as a monitored parameter during signal reconstruction. Consequently, research focuses on maximizing photon count rates by progressively reducing detector dead time, shifting the dominant performance bottleneck to the timing processing chain, which ultimately limits the maximum achievable data throughput. To this aim, we present a discrete-component, event-driven, four-channel shift register designed to enhance the photon temporal handling capabilities of current commercial state-of-the-art TCSPC instruments by leveraging their multichannel architectures. The proposed implementation operates at channel-switching frequencies up to 3.17 GHz, accepts input pulses with a minimum full width at half maximum (FWHM) of 148 ps, and introduces an intrinsic timing jitter contribution below 2.6 ps rms. Seamless integration into established measurement systems enables up to a fourfold increase in the sustainable per-channel count rate, while reducing the minimum admissible input pulse width.

Read PDF