Leadless Atrial Pacing Targeting Bachmann Bundle for Atrial Resynchronization: A Single Center Registry.

Oct 2026 · Heart Rhythm · 0 citations · 25 references
Medicine

TL;DR

The BBLAP implant technique presented in this registry is feasible and reduced IACD and further study on the effect of BBLAP on AF burden is needed.

Abstract

Background

Interatrial conduction delay (IACD) resulting from fibrosis or functional block in Bachmann bundle (BB) occurs often in patients with sinus node dysfunction (SND) and increases the risk of atrial fibrillation (AF). Pacing the BB in patients with SND may reduce AF burden. Little is known of the feasibility and effectiveness of BB leadless atrial pacemaker (BBLAP) implantation.

Objective

To determine if BBLAP is safe, feasible and improves IACD.

Methods

This registry prospectively enrolled 21 consecutive adults with baseline P wave duration (PWD) >120ms undergoing BBLAP for SND. Fluoroscopy, intracardiac echocardiography and left atrial conduction time (LACT) measurements from a decapolar coronary sinus (CS) catheter were used to facilitate implantation. BBLAP implant success was defined as improved PWD, P wave morphology, and LACT. Baseline and BBLAP LACT and PWD, axis, and morphology were compared.

Results

Of the 21 patients enrolled, 17 had successful BBLAP (81%). Among those with successful BBLAP, baseline PWD was 150 (±21) ms, BBLAP PWD was 113 (±23) ms, a reduction of 37 (±13) ms, p < 0.001. BBLAP P wave axis was inferiorly directed in all cases. 8/8 patients with a negative terminal P wave in the inferior leads at baseline had correction with BBLAP. LACT decreased from 113 (±35) ms to 78 (±33) ms, mean difference 35 (±15) ms, p <0.001, with BBLAP. 13/17 patients demonstrated a shift in CS activation from proximal-distal activation to mid- or distal-proximal activation during BBLAP. In all cases there were no complications.

Conclusion

The BBLAP implant technique presented in this registry is feasible and reduced IACD. Further study on the effect of BBLAP on AF burden is needed.

View source

Similar papers

Open access Oct 2026

Bachmann’s bundle vs conventional right atrial pacing: clinical and arrhythmic outcomes

Conventional right atrial (RA) pacing is associated with P-wave prolongation, interatrial conduction delay, and left-sided atrioventricular dyssynchrony, potentially leading to electromechanical dysfunction and increased arrhythmogenesis. Bachmann’s bundle (BB) pacing preserves physiological conduction and has be...

I. Cruz, M. Primo, D. Fernandes et al. · 0 citations
Open access Oct 2026

Biventricular vs conduction system pacing after atrioventricular node ablation for heart failure with atrial fibrillation: the CONDUCT-AF trial.

In HF patients with refractory AF undergoing atrioventricular node ablation, CSP with left bundle branch area pacing was non-inferior to BVP for LVEF improvement at 6 months, providing comparable clinical outcomes and a favourable procedural profile.

D. Žižek, M. Ivanovski, M. Mrak et al. · 0 citations
Open access Oct 2026

Atrial synchronization during bachmann bundle Pacing can be non-Invasively evaluated using lead V8

Bachmann bundle pacing (BBP) is an atrial pacing strategy that targets the dominant interatrial conduction pathway to synchronize interatrial conduction by advancing left atrial (LA) activation. However, the degree of interatrial synchronization remains difficult to quantify, and activation criteria are lacking....

S. Mulier, A. Fennema, V. Ramalingam et al. · 0 citations
Open access Sep 2026

Non-invasive assessment of left atrial activation using lead V8 during Bachmann bundle pacing

Abstract Aims Bachmann bundle pacing (BBP) improves atrial synchrony by advancing LA activation, but quantifying inter-atrial synchronization remains challenging. This study evaluated whether lead V8 reflects posterolateral LA activation time and whether lead V8 can be used to assess the effect of BBP on LA activation....

S. Mulier, A. Fennema, J. van Koll et al. · 0 citations
Open access Oct 2026

Conduction System Pacing for Resynchronization Therapy in Heart Failure: A Meta-Analysis of Randomized Controlled Trials.

BACKGROUND Conduction system pacing (CSP) has emerged as a promising modality for cardiac resynchronization therapy, but evidence of CSP in heart failure (HF) remains controversial. OBJECTIVES The authors aimed to compare the efficacy and safety of CSP, including left bundle branch area pacing (LBBAP) and His bundle...

P. Pajareya, W. Chuanchai, K. Laohasurayotin et al. · 0 citations
Open access Oct 2026

Long-term outcomes of single atrial lead pacing in sick sinus syndrome: risk stratification using wenckebach block rate

In patients with sick sinus syndrome (SSS), concerns about long-term atrioventricular (AV) conduction stability have led to frequent dual-chamber implantation. Although single atrial lead pacing (AAI pacing) is considered safe, prior studies were limited by short follow-up and prophylactic ventricular lead use, l...

A. Takami, M. Kato, F. Hirano et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.