Feasibility study of dose calculation based on contrast-enhanced computed tomography for stereotactic body radiation therapy of lung metastases treated with MR-Linac
When clinical risks are appropriately managed, CHCT has the potential to serve as a reference image for lung tumor SBRT on MR-Linac to optimize imaging utilization and enhance clinical workflow efficiency.
Abstract
The reliability of contrast-enhanced CT (CHCT) as a reference image for MR-based dose calculation in lung stereotactic body radiation therapy (SBRT) with MR-Linac remains uncertain due to contrast-induced electron density (ED) discrepancy. Whether CHCT can be used as a reference image for dose calculation for patients with lung tumors treated with MR-Linac still needs to be explored.
This retrospective study included 30 patients with lung metastases treated with SBRT (50 Gy in 5 fractions) on a 1.5 T MR-Linac. Non-contrast CT (pCT) and corresponding CHCT acquired in the same respiratory phase were used to extract mean electron density for targets and organs at risk (OARs). The treatment plans (TP
csCT
and TP
psCT
) based on synthetic CT (sCT) generated by assigning mean ED derived from pCT and CHCT were recalculated without altering parameters. Dosimetric discrepancies were evaluated using dose-volume histogram metrics, conformity indices, gradient indices and gamma analysis.
Significant differences in mean ED between pCT and CHCT were observed for several structures, most notably the lungs (mean differences > 12%), heart, aorta, and bronchial tree (
p
< 0.05), whereas minimal changes were found in bone and spinal cord. Despite these ED variations, dosimetric differences in target coverage were modest, with average discrepancy of 2.21% and 2.31% for TP
csCT
and TP
psCT
, respectively. Differences in the other dosimetric parameters (D
max
, D
2%
, D
95%
, D
98%
, D
mean
, CI and GI) for the target were generally below 3%. Although statistically significant differences were detected in some OARs dose metrics, all values remained well within clinical tolerance limits (3%).
Despite significant contrast-induced variations in ED, the use of CHCT as a reference image resulted in only minor dosimetric differences in lung SBRT. When clinical risks are appropriately managed, CHCT has the potential to serve as a reference image for lung tumor SBRT on MR-Linac to optimize imaging utilization and enhance clinical workflow efficiency.
HyperSight has the potential to be an alternative image used for dose calculation directly on HS-CBCT in the ART workflow, particularly in prostate cancer, and demonstrates good image quality and CT number accuracy.
INTRODUCTION
Numerous dosimetric studies have reported clinically insignificant differences between iridium-192 and cobalt-60. However, such comparison studies were conducted using TG-43 algorithms where heterogeneity is not accounted. Model based dose calculation algorithms (MBDCA) significantly improve dosimetric acc...
PURPOSE
Simulation-free radiation therapy (SFRT) using diagnostic computed tomography (dCT) may improve timeliness and reduce treatment burden in palliative care. We evaluated agreement between planned and delivered dose for nonadaptive dCT-based volumetric modulated arc therapy (VMAT) in palliative bone radiation ther...
M. O'Neil, Wendy Wells, J. Snir et al.· Practical Radiation Oncology· 0 citations
Purpose The benefit of proton beam therapy (PBT) for primary brain tumors compared with x-ray therapy has not been demonstrated by randomized controlled trials. We performed a model-based approach study to evaluate the potential benefit of PBT for brain tumors using the normal tissue complication probability (NTCP) mod...
H. Higaki, K. Nishioka, K. Yokokawa et al.· Advances in Radiation Oncolo...· 0 citations
Objectives To evaluate and compare conventional radiation dose metrics (computed tomography dose index volume [CTDIvol] and dose length product [DLP]) with size-specific dose estimates (SSDEs) in adult patients undergoing non-contrast computed tomography (NCCT) abdomen examinations. Materials and Methods This study aim...
S. Verma, Rohit Sharma, A. Mishra et al.· Indian Journal of Radiology...· 0 citations
We use cookies to run the site and, with your consent, for analytics and to show ads.
See our Cookie Policy.