Impact of Photon-Counting CT Acquisition Modes on Coronary Stenosis Assessment and Radiation Dose: A Phantom and Clinical Evaluation.

Oct 2026 · Zeitschrift für Medizinische Physik · 0 citations · 27 references
Medicine

TL;DR

While UHR acquisition provides highest image quality in PCD-CT cCTA, lower-dose acquisition modes achieve comparable diagnostic accuracy for coronary stenosis detection and Protocol selection should therefore be individualized, and prospective controlled studies are required.

Abstract

INTRODUCTION Photon-counting detector computed tomography (PCD-CT) offers high spatial resolution and dose-efficient coronary CT angiography (cCTA), but protocol selection, especially optimal acquisition mode, requires balancing spacial resolution, temporal robustness and radiation exposure. This study evaluated various PCD-CT acquisition modes in terms of image quality, blooming, stenosis grading, diagnostic accuracy, and radiation dose using phantom and clinical cCTA.

Methods

First-generation dual-source PCD-CT was used with four acquisition modes: high-pitch spiral with prospective ECG-gating (Flash), Sequence, Spiral, and retrospective ultra-high-resolution (UHR) spiral mode. Phantom experiments used six 3D-printed coronary artery models with stenoses ranging from 21% to 73%. In a retrospective clinical study, 41 cCTA examinations were analyzed. Qualitative and quantitative image assessments used invasive coronary angiography as reference standard for diagnostic accuracy. Patient characteristics and radiation doses were recorded.

Results

UHR demonstrated the highest image quality and the lowest quantitative blooming in phantom and clinical data (clinical image quality score: UHR 4.12 ± 0.99 vs. Flash 3.00 ± 0.86; p = 0.002). Diagnostic accuracy varied across protocols for detecting ≥50% and ≥70% stenoses was high across all modes (e.g., Flash: 76.5% each; Sequence: 100% and 57.1%; UHR: 64.7% and 82.4%). Radiation dose was highest in UHR (26.44 ± 9.76 mSv) compared with Sequence (13.65 ± 8.86 mSv) and Flash (1.40 ± 0.32 mSv); all p ≤ 0.011.

Conclusion

While UHR acquisition provides highest image quality in PCD-CT cCTA, lower-dose acquisition modes achieve comparable diagnostic accuracy for coronary stenosis detection. Protocol selection should therefore be individualized, and prospective controlled studies are required.

Read PDF

Similar papers

Open access Sep 2026

Dual-Source Photon-Counting CT for Thoracic Arteries: Enhancing Image Quality and Diagnostic Assessability with Low-Energy Virtual Monoenergetic Imaging

Background/Objectives: This study aimed to investigate the impact of low-energy virtual monoenergetic imaging (VMI) on quantitative and qualitative image characteristics, as well as its effect on the diagnostic assessability of thoracic arteries in photon-counting computed tomography angiography (CTA). Methods: We retr...

Andreea I. Nica, C. Booz, G. Bucolo et al. · 0 citations
Sep 2026

Diagnostic accuracy of photon-counting detector CT for detection of coronary in-stent restenosis.

BACKGROUND Photon-counting detector computed tomography (PCD-CT) offers improved spatial resolution and reduced blooming artifacts compared with conventional CT, potentially improving visualization of coronary stents. We evaluated the diagnostic performance of PCD-CT for detecting significant coronary in-stent restenos...

K. Aramouni, C. Kharsa, Patrick Sarkis et al. · 0 citations
Open access Aug 2026

Radiation and Contrast Medium Dosage Optimization of Coronary CT Angiography While Combining Personalized Patient Protocol Technology and Automated Tube Voltage Selection Technology

Objective: This study aims to investigate whether combining automated tube voltage selection (ATVS) with P3T for contrast injection in coronary computed tomography angiography (CCTA) allows for patient-tailored scanning protocols. Specifically, we assessed whether this combination maintains equivalent image quality and...

Jing-Kun Sun, Yang Zhou, Ge Zhang et al. · 0 citations
Open access Oct 2026

First clinical experience with the NAEOTOM Alpha.Peak photon-counting CT: high diagnostic reliability and ultra-low-dose CCTA in routine practice

The NAEOTOM Alpha.Peak demonstrated excellent image quality at remarkably low radiation doses in routine clinical use, and even during the early implementation phase, the system achieved high diagnostic reliability and efficiency in CCTA.

P. Breitbart, E. Yikit, E. Giokoglu et al. · 0 citations
Open access Sep 2026

Single-Bolus Sequential Triple-Rule-Out CT Angiography: Image Quality and Radiation Dose on Wide-Area Detector and Dual-Source CT

Background/Objectives: Single-pass triple-rule-out computed tomography (CT) angiography (TRO-CTA) must compromise among differing pulmonary, coronary, and aortic contrast-transit times. Wide-area detector CT (WAD-CT) and dual-source CT (DSCT) offer different coverage, temporal resolution, and dose profiles, but direct...

Hyunjung Kim, Jin Woo Kim, Sung-Jin Cha et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.