Oct 2026· Zeitschrift für Medizinische Physik· 0 citations· 27 references
Medicine
TL;DR
While UHR acquisition provides highest image quality in PCD-CT cCTA, lower-dose acquisition modes achieve comparable diagnostic accuracy for coronary stenosis detection and Protocol selection should therefore be individualized, and prospective controlled studies are required.
Abstract
INTRODUCTION
Photon-counting detector computed tomography (PCD-CT) offers high spatial resolution and dose-efficient coronary CT angiography (cCTA), but protocol selection, especially optimal acquisition mode, requires balancing spacial resolution, temporal robustness and radiation exposure. This study evaluated various PCD-CT acquisition modes in terms of image quality, blooming, stenosis grading, diagnostic accuracy, and radiation dose using phantom and clinical cCTA.
Methods
First-generation dual-source PCD-CT was used with four acquisition modes: high-pitch spiral with prospective ECG-gating (Flash), Sequence, Spiral, and retrospective ultra-high-resolution (UHR) spiral mode. Phantom experiments used six 3D-printed coronary artery models with stenoses ranging from 21% to 73%. In a retrospective clinical study, 41 cCTA examinations were analyzed. Qualitative and quantitative image assessments used invasive coronary angiography as reference standard for diagnostic accuracy. Patient characteristics and radiation doses were recorded.
Results
UHR demonstrated the highest image quality and the lowest quantitative blooming in phantom and clinical data (clinical image quality score: UHR 4.12 ± 0.99 vs. Flash 3.00 ± 0.86; p = 0.002). Diagnostic accuracy varied across protocols for detecting ≥50% and ≥70% stenoses was high across all modes (e.g., Flash: 76.5% each; Sequence: 100% and 57.1%; UHR: 64.7% and 82.4%). Radiation dose was highest in UHR (26.44 ± 9.76 mSv) compared with Sequence (13.65 ± 8.86 mSv) and Flash (1.40 ± 0.32 mSv); all p ≤ 0.011.
Conclusion
While UHR acquisition provides highest image quality in PCD-CT cCTA, lower-dose acquisition modes achieve comparable diagnostic accuracy for coronary stenosis detection. Protocol selection should therefore be individualized, and prospective controlled studies are required.
Background/Objectives: This study aimed to investigate the impact of low-energy virtual monoenergetic imaging (VMI) on quantitative and qualitative image characteristics, as well as its effect on the diagnostic assessability of thoracic arteries in photon-counting computed tomography angiography (CTA). Methods: We retr...
Andreea I. Nica, C. Booz, G. Bucolo et al.· Diagnostics· 0 citations
Objective: This study aims to investigate whether combining automated tube voltage selection (ATVS) with P3T for contrast injection in coronary computed tomography angiography (CCTA) allows for patient-tailored scanning protocols. Specifically, we assessed whether this combination maintains equivalent image quality and...
Jing-Kun Sun, Yang Zhou, Ge Zhang et al.· Diagnostics· 0 citations
The NAEOTOM Alpha.Peak demonstrated excellent image quality at remarkably low radiation doses in routine clinical use, and even during the early implementation phase, the system achieved high diagnostic reliability and efficiency in CCTA.
P. Breitbart, E. Yikit, E. Giokoglu et al.· European Heart Journal· 0 citations