Material-dependent effects of x-ray pre-filtration on material decomposition in photon-counting CT

Aug 2026 · Physics in Medicine and Biology · Vol 71 · 0 citations
Medicine Physics

TL;DR

These results demonstrate that optimal filtration strategies for PCCT are material dependent, and provide guidance for filter selection at PCCT MD applications, particularly those involving tantalum contrast agents, dual-contrast protocols pairing low and high-Z materials, or high-energy virtual monoenergetic image reconstructions for improved visualization of large body habitus.

Abstract

Objective. Characterize the influence of x-ray source filtration on material decomposition (MD) performance in photon-counting computed tomography (PCCT), understanding effects across different filter materials and thicknesses using contrast agents spanning low-Z (iodine, barium) and high-Z (tantalum) compositions. Approach. Single-contrast phantoms containing iodine, barium, or tantalum, and a dual-contrast phantom containing iodine and tantalum, were scanned on a CdZnTe PCCT system using several pre-filtration conditions: no pre-filter, 0.375 mm brass (single-contrast), and no pre-filter, 0.3 mm Cu, 2.1 mm Cu, 0.15 mm Sn, and 0.6 mm Sn (dual-contrast). MD basis map performance was quantified using voxel-level sensitivity, specificity, quantification accuracy, and contrast-to-pooled-noise ratio. Visual assessments (2D/3D axial slices, histograms) provided complementary analysis of material identification and quantification (MIQ). Main Results. Pre-filtration produced material dependent effects. For iodine and barium, brass filtration improved sensitivity at low concentrations, but increased crosstalk with PMMA and water, reducing specificity. For tantalum, brass filtration improved sensitivity (up to >60-fold increase) and quantification accuracy with increased crosstalk. In dual-contrast imaging, thin and thick Cu and thick Sn improved tantalum sensitivity and quantification, whereas iodine sensitivity and quantification remained stable. Thick Cu and Sn severely reduced specificity for both elements. Across experiments, linear regressions of measured versus ground truth concentrations remained linear (R2 >0.965) and confirmed detector response. Significance. These results demonstrate that optimal filtration strategies for PCCT are material dependent. Moderate filtration enhances MIQ for high-Z agents such as tantalum while preserving specificity, whereas filtration can degrade specificity for low-Z agents. The findings provide guidance for filter selection at PCCT MD applications, particularly those involving tantalum contrast agents, dual-contrast protocols pairing low and high-Z materials, or high-energy virtual monoenergetic image reconstructions for improved visualization of large body habitus. This work establishes a foundation for filtration PCCT protocol design and supports clinical translation of high-Z contrast agents.

View source

Similar papers

Open access Oct 2026

Scatter-induced non-linear signal compression and SDNR loss in triple-energy photon-counting iodine imaging.

Triple-energy x-ray imaging with photon-counting detectors (PCDs) enables material decomposition, but energy-dependent scatter introduces bin-specific distortions that propagate nonlinearly through triple-energy subtraction and degrade image quality. The purpose of this work is to quantify how bin-dependent scatter alt...

Jesse Tanguay, K. Sims · 0 citations

Physics in Medicine & Biology

This study demonstrates the feasibility of DE CBCT technology on the HyperSight imaging system through a novel statistical material specific imaging algorithm and VMI-based CNR improvements and metal artifact reduction.

Bryanna Smith, M. Sattarivand · 0 citations
Open access Sep 2026

Influence of X-ray Tube Voltage on Computed Tomography Values and Contrast-Noise Ratios in Virtual Monoenergetic Images in routine clinical Photon Counting Detector Computed Tomography: Does kV Still Matter?

This study investigated the kV dependence of VMI CT values in PCD-CT across a wide range of tissue types, keV levels, and kV settings for various phantom sizes, and to determine which kV settings yield the optimal contrast-to-noise ratio (CNR) for a given VMI energy and phantom size.

C. Jeukens, Floor Pinckaers, B. Martens et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.