B-035 Validation of Neurofilament Light Chain as laboratory developed test using electrochemiluminescence assay in serum and plasma
Abstract
Neurofilament light chain (NfL) is a 68 kDa type IV intermediate filament mainly found in axons and has structural functions mainly in the neuronal cytoskeleton. During axonal damage, NfL is released into the cerebrospinal fluid, resulting in increased concentrations in the blood as well. Increased NfL concentrations in both CSF and blood have been reported among individuals with traumatic brain injury, amyotrophic lateral sclerosis, multiple sclerosis, alzheimer disease and other neurodegenerative diseases. The objective of this validation was to evaluate the performance of Roche Elecsys NfL research use only (RUO) kits in serum and plasma as a laboratory developed test (LDT) for clinical trials testing per Clinical and Laboratory Standards Institute (CLSI) guidelines. NfL was measured quantitatively by two-incubation step assay using sandwich assay principle method on the Roche Cobas 8000 (Roche Diagnostics, USA) using Roche Elecsys RUO immunoassay kits. Intra-assay precision (3 levels, 20 replicates each) average CV was 2.9%. Inter-assay precision (3 levels, 1 replicate per day, 10 days) average CV was 4.7%. Accuracy was tested in 3 levels of kit controls and 3 different concentrations of NfL reference material (Roche NFLCore, 168 µg/mL) measured in replicates of 4. All samples were within manufacturer range, and average recovery for kit controls was 97.3%. Recoveries for NfL reference material ranged from 97.4 to 108.1%. Inter-lab comparison performed between two sites (N=58) resulted in a slope of 0.970, intercept 1.5, and correlation coefficient, (r=0.9997). Serum versus plasma correlation (N=31) resulted in a slope of 1.043, intercept 0.01, and correlation coefficient (r=0.9910). Analytical Measurement Range (0.3 to 5000 pg/mL) was verified using 5 levels of serum spiked with linearity material, 3 replicates each (slope = 0.991). Sensitivity (LLOQ) of 0.3 pg/mL was verified using 5 levels of serum spiked with linearity material, 8 replicates each per day, 5 days total 40 replicates per level (CV = 8.4%). Age specific reference ranges were established as: 4 years or younger= 1.56, 5 to 9 years= 1.30,10 to 14 years = 1.13, 15 to 19 years = 1.26, 20 to 29 years = 1.30, 30 to 39 years = 1.49, 40 to 49 years= 1.69, 50 to 59 years = 2.99, 60 to 69 years = 3.65, 70 to 79 years = 6.04, and 80 years or older = 9.13. Stability was evaluated at room temperature (20–25°C):14 days, refrigerated (2–8°C):14 days, frozen (< –10°C): 14 days, freeze/thaw cycles: up to 3 cycles, deep frozen (< –70°C): 199 days. Analytical interferences were evaluated for hemoglobin 1000 mg/dL, triglycerides 1500 mg/dL, conjugated bilirubin 40 mg/dL, unconjugated bilirubin 40 mg/dL. The observed bias for analytical interference parameters was =±10.0%. The laboratory developed test for NfL provides a robust quantitative method for determination of serum NfL and meets the Clinical and Laboratory Standards Institute (CLSI) requirements. These findings support its potential utility as a biomarker in therapeutic and drug development studies.