Sep 2026· Journal of refractive surgery· Vol 42 9, pp.
e868-e877
· 0 citations· 19 references
Medicine
TL;DR
In eyes undergoing cataract surgery, the Eye-star demonstrated superior repeatability to the IOLMaster for keratometry but notably inferior repeatability for most en face variables.
Abstract
Purpose
To compare the repeatability of and describe the agreement between the IOLMaster 700 (Carl Zeiss Meditec AG) and Eyestar 900 (Haag-Streit) under identical test conditions.
Methods
This was a comparative cross-sectional study conducted at a single-surgeon ophthalmology clinic in Australia. Adults presenting for cataract surgery underwent two scans each on both the IOLMaster and Eyestar in random order. Bland and Altman's methods were used to assess repeatability and agreement.
Results
A total of 160 left eyes satisfied inclusion. For anterior keratometry, the 95% repeatability (test-retest) errors for the Thibos power components were lower with the Eyestar (M = 0.23, J0 = 0.20, J45 = 0.15 diopters [D]) than the IOLMaster (0.28, 0.21, and 0.15 D). For posterior keratometry, the errors were also lower with the Eyestar (M = 0.05, J0 = 0.02, J45 = 0.03 D) than the IOLMaster (0.09, 0.07, 0.07 D). The segmental axial variables had comparable repeatability. White-to-white distance was more repeatable with the IOLMaster (0.14 mm) than the Eyestar (0.26 mm). The Eyestar had notably inferior repeatability for the chord of angle alpha (ICX = 0.51, ICY = 0.57 mm) compared to the IOLMaster (0.14, 0.24 mm). For the chord of angle kappa, the repeatability of the Eyestar (PCX = 0.48, PCY = 0.49 mm) was also inferior to that of the IOLMaster (0.17, 0.15 mm). The Eyestar consistently measured a lower posterior corneal power; the mean difference in the posterior flat curvature (PK1) was -0.40 D (95% CI: -0.42 to 0.40). The biometers had close agreement for the segmental axial variables.
Conclusions
In eyes undergoing cataract surgery, the Eye-star demonstrated superior repeatability to the IOLMaster for keratometry but notably inferior repeatability for most en face variables.
PURPOSE
To evaluate the repeatability of biometric measurements obtained with the MS-39 AL (CSO) and to assess their agreement with those provided by the IOLMaster 700 (Carl Zeiss).
SETTING
IRCCS Bietti Foundation, Rome, Italy.
DESIGN
Prospective evaluation of diagnostic test.
METHODS
Patients were classified in...
Gianluca Cappuccini, K. Hoffer, Arianna Grendele et al.· Journal of cataract and refr...· 0 citations
The three devices showed acceptable agreement at the population level, and notably, the average magnitude of disagreement was limited, at the individual level, however, differences exceeded clinically relevant thresholds in a meaningful proportion of patients, including some toric IOL candidates.
Purpose: To compare ocular biometry measurements obtained using optical low-coherence reflectometry (OLCR), swept-source optical coherence tomography (SS-OCT), and A-scan devices in eyes with cataracts and to evaluate prediction error based on the difference between the predicted refraction and the postoperative spheri...
Xiu-Ying Quan, Geonju Sakong, Kyong-Hwan Cho· Annals of optometry and cont...· 0 citations
Purpose: To evaluate the feasibility of contrast sensitivity (CS) testing using a head-mounted binocular perimeter (IMO vifa) and to describe postoperative CS in eyes implanted with three types of intraocular lenses (IOLs). Patients and Methods: This retrospective single-center study included 102 eyes of 71 patients wh...
Tetsutaro Oki, Koji Komatsu, Yuri Shinohe et al.· Clinical Ophthalmology· 0 citations
Objectives: To assess the repeatability and reliability of anterior and posterior corneal measurements in keratoconus patients who underwent corneal allogenic intrastromal ring segment (CAIRS) implantation using a Scheimpflug-based tomography system. Methods: This prospective observational study included 42 eyes: 20 ke...
Uğur Tunç, Dilan Çolak, Burcu Yakut et al.· Beyoglu Eye Journal· 0 citations
Purpose: This study aimed to compare the repeatability and agreement of corneal tomographic measurements between the Anterion Cornea App and Pentacam HR in keratoconus (KCN). Methods: Eyes with KCN were imaged three consecutive times with each device. Intra-device repeatability was assessed using intraclass correlation...
Ramadhan Ahmed, K. Reddy, Mahad Mohamed et al.· Bioengineering· 0 citations
We use cookies to run the site and, with your consent, for analytics and to show ads.
See our Cookie Policy.