The Challenge of Rendering Abnormalities Accurately Using Deep Learning-Based Reconstruction in Accelerated Brain MRI: A Disproportionate Loss of Clinically Relevant Image Information at Higher Rates of Acceleration.
A disproportionate, pathology-localized degradation of reconstruction quality as acceleration increases in DL-reconstructed brain MRI in normal and abnormal cases is demonstrated, highlighting the need for reconstruction strategies tailored to maintaining diagnostically critical features.
Abstract
Objectives
Deep learning (DL)-based reconstruction methods can substantially accelerate MRI acquisition while preserving image quality relative to traditional techniques. However, concerns remain regarding their ability to accurately reconstruct abnormalities at higher acceleration factors. This study quantitatively assesses the impact of higher-order acceleration on image quality in DL-reconstructed brain MRI in normal and abnormal cases, emphasizing the utmost need to render abnormal findings accurately.
Materials And Methods
Raw k-space image data from 5847 brain MRI examinations (fastMRI) were used. A radiologist-annotated subset of 1001 examinations (fastMRI+) served as the test set; the remainder were split into training (n=3715) and validation (n=1131) sets. Images were retrospectively undersampled (acceleration factors 2 to 12) and reconstructed using an end-to-end variational network. Image quality metrics, including normalized mean squared error (NMSE), structural similarity index measure (SSIM), and peak signal-to-noise ratio (PSNR), were computed within bounding-box annotations containing abnormalities and in size-matched normal control regions, and at the slice-level. The region-by-acceleration interaction was tested using linear mixed-effects models with a subject-level random intercept.
Results
All metrics demonstrated progressive degradation with increasing acceleration across slice-level and bounding-box-level analyses. At lower acceleration rates, bounding boxes with and without abnormalities showed comparable image quality. However, above 4-fold acceleration, NMSE within abnormal regions exhibited a notably sharper increase with rising acceleration compared to normal regions, while abnormal regions demonstrated a significantly faster decline in PSNR and SSIM. This nonsymmetric information loss was particularly evident for NMSE at higher acceleration rates. The region-by-acceleration interaction was statistically significant for all 3 metrics at both levels (all adjusted P<0.001), with the effect largest and most consistent for NMSE.
Conclusions
Our findings demonstrate a disproportionate, pathology-localized degradation of reconstruction quality as acceleration increases. This emphasizes the importance of rigorous clinical assessment of image quality in DL-MRI reconstruction, particularly at higher undersampling rates, highlighting the need for reconstruction strategies tailored to maintaining diagnostically critical features.
A comprehensive comparative evaluation of six U-Net variants, including U-Net, Attention U-Net, Residual U-Net, Recurrent-Residual (R2) U-Net, Residual-Attention U-Net (RAU-Net), and U-Net++, for image domain reconstruction of low-field MRI data from the M4RAW dataset is presented.
Akif Ahmed Nasif Purno, K. M. T. K. Siddiki, S. M. Chapal Hossain· Magnetic Resonance Imaging· 0 citations
Obtaining high-resolution brain MRI in small animals is challenging due to the inherent trade-off between image quality and acquisition time. Deep learning-based reconstruction (DLR) has emerged as a solution to improve image quality without prolonging scan time; however, its clinical utility and potential risks, such...
Wooseok Jin, Hye-Ran Na, Sang-Kwon Lee et al.· Veterinary Radiology & Ultra...· 0 citations
DLIR combined with high-frequency kernels reduced image noise while preserving spatial resolution and FBP-like noise texture, outperforming IR in detectability for most lesion sizes, concentrations, and dose levels.
A. Viry, P. Monnin, C. Pozzessere et al.· European Radiology Experimen...· 0 citations
Purpose To determine whether structural MRI information can improve reconstruction of highly accelerated 3D snapshot chemical exchange saturation transfer (CEST) MRI. Methods Fully sampled brain CEST data were retrospectively undersampled at acceleration factors (AFs) of 4, 6, 8, 10, and 12. We compared reconstruction...
E. Mensah, Abrar Faiyaz, G. Schifitto et al.· bioRxiv· 0 citations
Magnetic resonance imaging (MRI) reconstruction is highly sensitive to undersampling and acquisition artifacts. Although deep learning-based approaches have significantly improved reconstruction quality, they are typically optimized using global fidelity metrics that may not fully reflect reconstruction quality within...
A. Bekbolat, Almira Askhatova, Aibek S. Niyetkaliyev et al.· MethodsX· 0 citations
We use cookies to run the site and, with your consent, for analytics and to show ads.
See our Cookie Policy.