Skip to content
Preprint

Performance Evaluation of a High-Granularity LYSO-SiPM-Based Position-Sensitive Detector for a One-Shot Gamma-Scanning System with Sub-Millimeter Spatial Resolution

Aug 2026 · 0 citations · 30 references
Physics

Abstract

A compact position-sensitive {\gamma}-detector based on a thin monolithic LYSO crystal and a 96-channel SiPM array is developed for the spatial characterization and calibration of segmented {\gamma}-ray detector systems used in nuclear-physics experiments. The detector is designed to provide localized irradiation and rapid two-dimensional response mapping. A 7 cm diameter and 3 mm thick LYSO crystal is optically coupled to the SiPM array, and the position of the incident {\gamma}-ray interaction is reconstructed from the relative scintillation-light signals collected by neighboring SiPM channels. An asymmetry-based charge-sharing method is employed to determine the interaction position from the spatial distribution of the detected scintillation light. Detailed GEANT4 simulations, including optical photon transport, were performed to investigate the detector response and estimate its intrinsic spatial resolution. The simulations predict a spatial resolution of approximately 0.5 mm for 60 keV and 511 keV {\gamma}-rays under idealized conditions. A prototype detector was developed and experimentally characterized using coincidence measurements with the {\gamma}-scanning facility at GSI, Germany. The experimental measurements demonstrate a spatial resolution better than 1 mm in the central detector region, while the position-dependent response and degradation near the detector boundaries are investigated. The experimental results are compared with GEANT4 predictions to identify the contributions of optical photon transport, charge sharing, and detector geometry to the measured spatial resolution. The developed detector provides a compact solution for the three-dimensional characterization of highly segmented {\gamma}-ray detector arrays in nuclear physics experiments.

View source

Similar papers

Open access Sep 2026

The Characterization of a Compact CLLB:Ce Scintillation Detector for Mixed-Field Monitoring at a Muon Facility

Compact scintillation detectors with simultaneous gamma and neutron sensitivity are increasingly relevant for mixed-field monitoring and nuclear measurements under size and cost constraints. This work characterizes a compact Cs2LiLaBr6:Ce (CLLB:Ce) detector with integrated SiPM readout at the ISIS Muon Facility. We ass...

Paola Monza, Sayani Biswas, Matteo Cataldo et al. · 2 citations
Conference Open access Oct 2026

Neutron Beam Imaging on DICER Using a 6Li-Glass Scintillator Read Out by a Large Area Picosecond Photodetector (LAPPD)

Accurate neutron beam profiling and position verification are required for alignment, improvement of signal-to-noise, and imaging development, yet are often performed with single-use beam-imaging media that provide only integrated exposure and limited real-time feedback. We present a first analysis of neutron imaging d...

D. Král, A. Stamatopoulos, A. Couture et al. · 0 citations
Open access Oct 2026

Detector Response and Energy-Threshold Optimization for Imaging with Cosmic-Ray Muon-Induced Secondary Radiation

Imaging with cosmic-ray muon-induced secondary radiation complements muon transmission radiography and muon scattering tomography by exploiting the electrons and bremsstrahlung photons created along the muon track inside the target. Because these secondaries are produced predominantly at low energies, the performance o...

K. Bikit, Gergő Hamar, D. Velimirović et al. · 0 citations
Oct 2026

Towards multiparticle position sensitivity with an opaque scintillation detector

LiquidO is a detection technique that utilizes opaque media to stochastically confine scintillation and Cherenkov light near its production point, resulting in effective optical segmentation. A lattice of wavelength-shifting (WLS) fibers suspended in the detector volume collects and transports light to photosensors at...

Sander Katz, G. Wendel, Jacob Moffatt et al. · 0 citations
Oct 2026

Gamma-ray response of a GAGG:Ce scintillator coupled to a 2 × 2 SiPM array

A relatively large 35 × 35 × 20 mm3 GAGG:Ce scintillator coupled to a 2 × 2 silicon photomultiplier (SiPM) array was characterized for gamma-ray detection, including relative-gain calibration, nonlinear energy response, energy resolution, and timing resolution. The SiPM-array response was well described by a saturation...

Zi-Bo Qin, Zai-Hong Yang, Kai-Jie Zhou et al. · 0 citations
Oct 2026

Design and performance evaluation of a GAGG:Ce long-bar Compton camera with dual-ended DOI readout using Geant4 simulations.

Compton cameras enable medium- and high-energy gamma-ray imaging without mechanical collimation. Extending an individual detector element increases the active volume without adding readout channels, but also increases axial position uncertainty and can degrade angular resolution. This study evaluated whether a cerium-d...

Jia-Feng Li, Ming Zhang, Yong-Li Mu et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.