Analysis of the application of computerized optokinetic nystagmus analyzer in vision examination

Aug 2026 · BMC Ophthalmology · Vol 26 · 0 citations · 21 references
Medicine

TL;DR

The computerized OKN analyzer exhibits favorable reliability and practicability for VA testing in patients aged aged ≥ 6 years across different refractive errors, with accuracy that appears sufficient for screening and general clinical estimation of VA.

Abstract

To investigate the practicability, accuracy, and consistency of the computerized optokinetic nystagmus (OKN) analyzer for visual acuity (VA) measurement. Cross-sectional consistency evaluation study. A total of 219 subjects (13.48 ± 12.89 years, 115 males, 104 females) underwent bilateral VA testing using a computerized OKN analyzer and a traditional OKN drum. Additionally, 104 subjects (12.58 ± 12.05 years, 54 males, 50 females) were tested bilaterally with the computerized OKN analyzer and a standard logarithmic E chart (random order), including 80 myopic eyes (spherical equivalent [SE] ≤ -0.5 D) and 63 hyperopic eyes (SE ≥ + 0.5 D); VA was recorded in logMAR. Concordance rates between the OKN analyzer and OKN drum were calculated. Intraclass correlation coefficients (ICC) and Bland-Altman plots were used to assess the consistency between the OKN analyzer and a standard logarithmic E chart. Subjects completing the lightbox test were stratified by age (3–4 years, 5 years, 6 years, 7–18 years, 18–58 years), refractive error, and VA range to compare inter-method correlations. Compared with the OKN drum, the computerized OKN analyzer achieved a concordance rate of 74.00% (right eye) and 77.08% (left eye), indicating good agreement. For VA measurements, mean VA (logMAR) of the right eye was 0.24 ± 0.29 (OKN analyzer) vs. 0.24 ± 0.28 (lightbox); left eye was 0.25 ± 0.28 vs. 0.24 ± 0.29. ICC values were 0.92 (right eye) and 0.88 (left eye, both P < 0.001), confirming significant consistency. Bland-Altman analysis showed 95% limits of agreement of -0.228 to 0.233 (right eye) and − 0.266 to 0.293 (left eye). Inter-method consistency improved with age (highest in ≥ 18 years) and was stronger in myopic vs. hyperopic eyes. Notably, correlation weakened as VA improved, with the highest concordance in the 0.1–0.5 decimal VA. The computerized OKN analyzer exhibits favorable reliability and practicability for VA testing in patients aged ≥ 6 years across different refractive errors, with accuracy that appears sufficient for screening and general clinical estimation of VA. However, in very young children (3–4 years) and in eyes with normal VA, agreement with chart‑based VA is limited, and further validation is required.

Read PDF

Similar papers

Open access Sep 2026

Objective Assessment of Intermittent Exotropia Magnitude and Control Using the Binocular Eye Alignment Monitor.

The notable differences in control observed with the BEAM compared to current clinical measures suggest that office control scores and prolonged objective monitoring may capture different aspects of IXT control.

Erin C. Jenewein, Christopher J. Morris, Mitchell Scheiman et al. · 0 citations
Open access Oct 2026

Comparison of Anterior Segment Biometric Parameters Measured by Pentacam and Sirius in Healthy Myopic Eyes.

Pentacam and Sirius measurements showed high correlation and good agreement and can provide complementary clinical references, however, systematic biases were observed in the absolute ACD, Kf, Ks, and AST values obtained with the two devices.

Han Song, Wei Yang, Xiao-Xue Wang et al. · 0 citations
Sep 2026

Comparing the Single Maddox Rod Test to the Alternate Cover Test in Normal Individuals.

PURPOSE The single Maddox rod test (SMRT) is a rapid subjective method for detecting ocular misalignment. In this study, we compared the prism single Maddox rod test (PSMRT) to the prism alternate cover test (PACT) in normal individuals to see if there was any disagreement. METHODS Ocular alignment was measured at di...

Victoria Erickson, K. Merrill, Collin M. McClelland et al. · 0 citations
Open access Sep 2026

Visual Acuity: Comparison of ETDRS Charts Using Latin Letters, Tumbling E, or Landolt C Optotypes

Non-letter optotypes provide clinically useful and broadly comparable measures of VA at the population level supporting their use in clinical practice or population/clinical studies and wide variability means they cannot be considered interchangeable for individual subjects.

Michael Kalloniatis, Henrietta Wang, Kimberly Meier et al. · 0 citations
Review Jul 2026

A Pilot Study to Compare a Portable Tablet Perimeter with Humphrey Field Analyzer in Patients with Neuro-Ophthalmic Disorders

ABSTRACT Purpose To evaluate the diagnostic accuracy of the Melbourne Rapid Fields (MRF) tablet perimeter in comparison with the Humphrey Visual Field (HVF) analyzer in detecting visual field defects in patients with neuro-ophthalmic conditions. Methods This prospective observational study was conducted on 15 patients...

Gunjan Saluja, Lavanya Gandepalli, V. Garg et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.