Accuracy, safety, and usability of a real‐time continuous glucose monitoring system in pediatric diabetes management

Aug 2026 · Journal of Diabetes Investigation · Vol 17, pp. 1807 - 1815 · 0 citations · 26 references
Medicine

TL;DR

The real‐time CGM system demonstrated high accuracy, safety, and usability in the glycemic monitoring of children and adolescents, and showed high safety, given that only 3 device‐related AEs in 2 participants were reported, and no serious adverse events (SAEs) were reported.

Abstract

ABSTRACT Background The purpose of the study was to evaluate the accuracy, safety, and usability of a real‐time Continuous Glucose Monitoring (CGM) system compared with venous blood glucose (vBG) in children and adolescents. Methods The study enrolled 81 participants aged from 3 to 17 at three sites in China. Two sensors (GS1, SiBionics, Shenzhen) were inserted on the back of each upper arm for up to 14 days for each participant. A variety of analyses were conducted for evaluating accuracy by comparing the CGM readings and vBG values, including 20%/20 agreement rate, the mean absolute relative difference (MARD), the Clarke error grid, and consensus error grid analyses. Safety was measured by adverse event (AE) monitoring. AEs were documented, and the incidence rate was calculated. Usability was evaluated by self‐reported questionnaires. Results Data from 80 participants were analyzed. Our results demonstrated high accuracy of the real‐time CGM system. The 20%/20 agreement rate across all glycemic ranges was 93.9%. The MARD value was 8.7%. The results of the Clarke and consensus error grid analyses in zone A + B were 99.6% and 100%, respectively. The average score of usability was 95.3 ± 7.6, reflecting high satisfaction. The CGM also showed high safety, given that only 3 device‐related AEs in 2 participants were reported , and no serious adverse events (SAEs) were reported. Conclusion The real‐time CGM system demonstrated high accuracy, safety, and usability in the glycemic monitoring of children and adolescents.

Read PDF

Similar papers

Open access Oct 2026

Performance of a 15-Day Continuous Glucose Monitoring System in Children and Adolescents With Diabetes

Background: The CareSens Air (CSAir) continuous glucose monitoring system is approved for use in adults in South Korea and received adults-only CE marking in early 2024. The aim of this prospective, multicenter, pivotal study in children and adolescents with insulin-dependent diabetes mellitus was to determine if the p...

Manuela Link, C. Denzer, T. Biester et al. · 0 citations
Review Open access Sep 2026

Usability, acceptability and feasibility of continuous glucose monitoring among children and adolescents with type 1 diabetes in Kenya

Introduction Continuous Glucose Monitoring (CGM) is considered standard care in high-income countries. There is, however, limited published evidence on CGM use in low- and middle-income countries. The purpose of this study was to assess the usability, acceptability, and feasibility of CGM use among people living with t...

P. Amolo, L. Mungai, A. K. Karume et al. · 0 citations
Open access Oct 2026

Accuracy and Safety of the LinX Continuous Glucose Monitoring System in Adults With Diabetes: A Prospective Multicentre Open‐Label Clinical Performance Study

Accurate continuous glucose monitoring (CGM) is essential for diabetes management. This study evaluated the accuracy and safety of the LinX CGM system over 15 days in adults with diabetes.

Yang Liu, Yu-Long Chen, Yi-Mei Chen et al. · 0 citations
Oct 2026

Pediatric Inpatient Continuous Glucose Monitoring: Development and Implementation of a Hospital Policy for Accuracy-Validated Insulin Dosing.

An inpatient CGM policy from an adult to pediatric hospital setting is translated, supporting high levels of protocol adherence, clinical accuracy, safety, and improved experiences for clinicians.

Ming-Yeh Lee, Hector I. Ortega, Christopher Chow-Parmer et al. · 0 citations
Open access Sep 2026

Pediatric Inpatient Continuous Glucose Monitor Accuracy by Sensor Wear Day, Hospital Day, and Glucose Rate of Change

In this pediatric hospital CGM accuracy study, MARD was within a favorable range of <14% overall and throughout sensor wear and hospitalization, highlighting the need to incorporate trend arrows into hospital CGM protocols.

Jessica L. Ruiz, A. Stein, Alexander Wheelock et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.