Skip to content
Open access

AI-Assisted Stability-Indicating RP - HPLC Method Development and Validation for the Quantitative Estimation of Tafamidis in Bulk Drug and Pharmaceutical Dosage Forms

Sep 2026 · Journal of Pharmaceutical Research International · 0 citations

Abstract

A stability-indicating reverse-phase high-performance liquid chromatographic (RP-HPLC) method was developed and validated for the quantitative estimation of Tafamidis in bulk drug and pharmaceutical dosage forms. Chromatographic separation was achieved using a Phenomenex Gemini ODS C18 column (250 × 4.6 mm, 5 µm) with methanol and phosphate buffer (25:75, v/v) as the mobile phase at a flow rate of 1.0 mL/min. Detection was performed at 312 nm using a PDA detector. The method was evaluated for system suitability, specificity, linearity, accuracy, precision, intermediate precision, robustness, limit of detection, limit of quantification, and stability-indicating capability. Tafamidis showed a mean retention time of approximately 3.16 min and linear response over the concentration range of 10–50 µg/mL, with a correlation coefficient of 0.999. The assay of the pharmaceutical dosage form was 99.23%. Recovery at the evaluated concentration levels was approximately 100.41–100.44%, while method precision and intermediate precision showed low percentage relative standard deviations. The calculated LOD and LOQ were 0.597 and 1.811 µg/mL, respectively. Forced-degradation studies under acidic, alkaline, oxidative, thermal, photolytic, and hydrolytic conditions demonstrated separation of Tafamidis from degradation products. The experimental dataset also provides a basis for future evaluation of artificial intelligence and machine-learning approaches for chromatographic optimisation; however, such models were not used to generate the reported validation results.

Read PDF

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.