Skip to content
Open access

Single Center Experience using Fiber Optic RealShape (FORS) Technology in Patients Treated with Complex Endovascular Aneurysm Repair.

Aug 2026 · Journal of Endovascular Therapy · pp. 15266028261479886 · 0 citations · 15 references
Medicine

TL;DR

Fiber Optic RealShape has an acceptable technical success rate in visceral TV catheterization during cEVAR, which has improved following the 3D-Hub implementation, supporting initial FORS use in all catheterizations to minimize radiation.

Abstract

Purpose

Fiber Optic RealShape (FORS) is a novel technology using light instead of X-rays to visualize guidewire and catheters during endovascular procedures. A 3D-Hub allows visualization of conventional catheters when used in combination with the FORS guidewire. This study evaluates the FORS performance and radiation exposure during visceral target vessel (TV) catheterization in complex endovascular aortic repair (cEVAR).

Materials And Methods

Single-centre data were retrieved from the prospective FORS Learn Registry cohort study. The primary outcome was FORS technical success, defined as successful catheterization using FORS without reverting to a conventional guidewire. Secondary outcomes included catheterization and fluoroscopy time, Dose Area Product (DAP) and Air Kerma (AK), compared across technical success, technical failure and conventional catheterizations using analysis of variance (ANOVA) with post hoc testing.

Results

Between December 2020 and December 2023, 96 patients underwent cEVAR. In total 319 visceral targets were catheterized, including 176 FORS attempts with a technical success rate of 63.6%. This success rate decreased in TVs with versus without ostial stenosis (21.7% vs 69.9%, P < .001) or with versus without calcification (14.3% vs 67.9%, P < .001), and increased with versus without the use of the 3D-Hub (80.6% vs 60.0%, P = .039). Type of TV, stent graft configuration or prior TV stent did not affect technical success rates. Radiation exposure was significantly lower during successful FORS catheterizations (fluoroscopy time 1.7 ± 2.1 min; DAP 3.7 ± 8.5 Gy∙cm2; AK 30.8 ± 55.9 mGy) compared to FORS technical failures (fluoroscopy time 6.5 ± 4.3 min, P < .001; DAP 8.8 ± 6.4 Gy∙cm2, P = .030; AK 126.5 ± 116.5 mGy, P < .001) and conventional catheterizations (fluoroscopy time 6.2 ± 9.2 min, P < .001; DAP 9.4 ± 20.25 Gy∙cm2, P = .003; AK 111.5 ± 151.1 mGy, P < .001). No significant differences occurred between technically failed and conventional catheterizations. Catheterization time was shortest with FORS technical success (6.5 ± 5.0 min) compared to conventional catheterizations (9.7 ± 13.5 min, P = .015), followed by technical failure (14.3 ± 6.9 min, P < .001).

Conclusion

Fiber Optic RealShape has an acceptable technical success rate in visceral TV catheterization during cEVAR, which has improved following the 3D-Hub implementation. Successful FORS catheterization results in reduced radiation exposure and catheterization time. When reversion to conventional catheterization is required, outcomes remain comparable to primary conventional procedures, supporting initial FORS use in all catheterizations to minimize radiation. Continued device and software development is essential to optimize FORS performance.Clinical ImpactFiber Optic RealShape (FORS) is a new technology that enables visualization of exclusive FORS-guidewires and catheters using light pulses instead of radiation. This study shows acceptable technical success rates with reduced radiation exposure compared to technically unsuccessful FORS and primary conventional catheterizations of visceral target vessels during complex endovascular aortic repair. These findings support the clinical benefit of initially attempting catheterization with FORS, as it results in reduced radiation exposure, even in case of reversion to conventional devices, without compromising patient outcomes. Additionally, the technical success rates improved significantly following the implementation of the 3D Hub.

Read PDF

Similar papers

Review Open access Aug 2026

Use of Drivewire 24 guidewire in neurointervention: initial clinical experience and technical advancements

A retrospective review of endovascular procedures at a center in which the Drivewire 24 (DW24) guidewire was used, from its introduction in the authors' department in May 2024 through January 2026, identified several unique technical applications of the DW24 that represent a meaningful advancement in neurointerventiona...

Uri Singfer, Ronette Goodluck Tyndall, J. Sousa et al. · 0 citations
Open access Aug 2026

The ATTAIN Technique (Amplatz lefT guiding caTheter pArtIal deploymeNt): An Alternative Approach for Transradial Stenting of Low-Origin Vertebral Artery Ostial Stenosis.

The ATTAIN technique showed preliminary feasibility and provided a standardized AL guiding catheter-assisted deployment sequence for selected right-sided low-origin VAOS, and provided a standardized AL guiding-catheter-assisted deployment sequence for selected right-sided low-origin VAOS.

Han-Ming Ge, Xiao-Bo Zhang, Zhan-Hua Liang et al. · 0 citations
Open access Aug 2026

Three-dimensional simulations for planning fenestrated/branched endografts in endovascular aneurysm repair for complex abdominal aortic aneurysm: a retrospective cohort study

F/B-EVAR guided by 3D simulation may allow for accurate planning of fenestration and branch positioning, potentially improving procedural success rate and suggesting a potential to enhance the quality of outcomes.

Yujing Zhang, Jing Wang, Cheng-Long Qu et al. · 0 citations
Open access Aug 2026

A Novel Electromagnetic Coil (EMAG) Fenestrated Stent-Graft for Rapid Cannulation to Reduce Use of Contrast Agent and Radiation Exposure During Fenestrated EVAR.

The findings of this study demonstrate the feasibility of the EMAG system as a novel approach for assisting branch-vessel cannulation during fenestrated endovascular aortic repair, offering a promising approach to improve patient safety in an increasingly common procedure.

Bryan W. Tillman, Gordon K Bryson, Catherine C. Go et al. · 0 citations
Open access Aug 2026

Construct Validity of a Novel Fiberoptic TAVR Guidewire for Invasive Pressure Measurements.

The fiberoptic guidewire effectively classified clinically relevant transvalvular gradients with limited bias post-TAVR, although it produced higher LVEDPs, possibly reflecting wire-induced leaflet interaction.

Geert A. A. Versteeg, M. Westerhof, N. Stens et al. · 0 citations